Separator Tubes Assessed for Measurement of 25-Hydroxyvitamin D3

By LabMedica International staff writers
Posted on 17 Feb 2016
Print article
Image: The Architect Immunoassay analyzer used for measuring 25-hydroxyvitamin D3 (25OHD3) (Photo courtesy of Abbott Diagnostics).
Image: The Architect Immunoassay analyzer used for measuring 25-hydroxyvitamin D3 (25OHD3) (Photo courtesy of Abbott Diagnostics).
Image: BD serum-separating advance vacutainer tubes (SST II) (Photo courtesy of Becton Dickinson).
Image: BD serum-separating advance vacutainer tubes (SST II) (Photo courtesy of Becton Dickinson).
The accuracy of 25-hydroxyvitamin D3 (25OHD3) measurement on specimens collected into serum separator tubes (SSTs) has been questioned because of possible interference by the gel.

High performance liquid chromatography (HPLC) and liquid chromatography-tandem mass spectrometry (LC-MS/MS) are still considered the reference methods for vitamin D measurement although commercial immunoassays have been developed to satisfy increased demand for 25OHD measurement.

A team of scientists including those from the Royal Surrey County Hospital and their colleagues in Saudi Arabia collected blood specimens were from 50 healthy volunteers (22 males, 28 females, aged 42.6 ± 11.2 and BMI 27.4 ± 5.2 kg/m2), ten of whom were taking 25OHD3 supplements. Blood specimens were simultaneously collected into plain vacutainer tubes and serum-separating advance vacutainer tubes (SST II) using a standard venesection procedure. Specimens were centrifuged within one hour of blood collection and the serum transferred into secondary tubes. Serum was stored at -80 °C until analysis six months later.

All vacutainers used were from Becton Dickinson (BD; Franklin Lakes, NJ, USA) and 25OHD3 was assayed on serum using high performance liquid chromatography (Waters Corporation; Milford, MA, USA), and immunoassays from Architect (Abbott Diagnostics; Abbott Park, IL, USA) and Liaison (DiaSorin; Saluggia, Italy).

The authors found there were no significant differences between 25OHD3 results (means ± SE, nmol/l) obtained from specimens collected into plain tubes and SSTs assayed by HPLC (39.0 ± 2.7 vs. 39.3 ± 2.7), Liaison (32.9 ± 2.2 vs. 32.8 ± 2.3), or Architect (43.1 ± 2.8 vs. 43.2 ± 2.8). In specimens collected into plain tubes and SSTs, 25OHD3 measurements by HPLC correlated significantly with those from the Architect and Liaison respectively.

The authors concluded that the gel in BD’s SSTs does not interfere with the measurement of 25OHD3 by HPLC or by the commercial immunoassays tested. This important finding may enable clinical laboratories to make cost savings by using SSTs without concerns about inaccuracy. The study was published in the January 2016 issue of the Journal of Clinical Laboratory Analysis.

Related Links:

Royal Surrey County Hospital
Becton Dickinson 
Waters Corporation 


New
Gold Member
Rotavirus Test
Rotavirus Test - 30003 – 30073
Verification Panels for Assay Development & QC
Seroconversion Panels
New
Silver Member
Total Hemoglobin Monitoring System
GREENCARE Hb
New
Bordetella Pertussis Molecular Assay
Alethia Pertussis

Print article

Channels

Molecular Diagnostics

view channel
Image: The study investigated D-dimer testing in patients who are at higher risk of pulmonary embolism (Photo courtesy of Adobe Stock)

D-Dimer Testing Can Identify Patients at Higher Risk of Pulmonary Embolism

Pulmonary embolism (PE) is a commonly suspected condition in emergency departments (EDs) and can be life-threatening if not diagnosed correctly. Achieving an accurate diagnosis is vital for providing effective... Read more

Immunology

view channel
Image: The findings were based on patients from the ADAURA clinical trial of the targeted therapy osimertinib for patients with NSCLC with EGFR-activated mutations (Photo courtesy of YSM Multimedia Team)

Post-Treatment Blood Test Could Inform Future Cancer Therapy Decisions

In the ongoing advancement of personalized medicine, a new study has provided evidence supporting the use of a tool that detects cancer-derived molecules in the blood of lung cancer patients years after... Read more

Microbiology

view channel
Image: Schematic representation illustrating the key findings of the study (Photo courtesy of UNIST)

Breakthrough Diagnostic Technology Identifies Bacterial Infections with Almost 100% Accuracy within Three Hours

Rapid and precise identification of pathogenic microbes in patient samples is essential for the effective treatment of acute infectious diseases, such as sepsis. The fluorescence in situ hybridization... Read more
Sekisui Diagnostics UK Ltd.