Performance Evaluated for Point-of-Care HbA1c Analyzer
|
By LabMedica International staff writers Posted on 07 Jul 2015 |
Glycated hemoglobin (HbA1c) levels, which reflect the average plasma glucose concentration over the preceding three months, play a pivotal role in the diagnosis, assessment, and monitoring of diabetes.
The immediate feedback of HbA1c levels is highly effective for controlling plasma glucose levels and point-of-care (POC) instruments, characterized as fast, portable, and easy-to-use, have been shown to be suitable for providing rapid feedback of HbA1c levels.
Medical laboratory scientists at the Seoul National University College of Medicine (Republic of Korea) assessed the analytical performance of a newly developed POC HbA1c analyzer. To investigate the accuracy of the analyzer, they measured the deviation from the International Federation of Clinical Chemistry (IFCC) reference targets using two sets of 12 duplicate fresh venous whole blood specimens with a range of 33.3 mmol/mol to 91.3 mmol/mol in International System of Units (SI) units, or 5.2% to 10.5% in National Glycohemoglobin Standardization Program (NSGP) units.
The SD A1cCare analyzer (SD Biosensor, Inc.; Suwon, Korea), which utilizes an immunochromatography assay, was used to measures total hemoglobin and HbA1c levels via its optical system and calculates the percentage HbA1c. The SD A1cCare Test kit uses an anti-HbA1c antibody that is specific for the first few amino acid residues of the glycated N-terminus of the β-chain of hemoglobin A0. The scientists tested 150 residual specimens using both the SD A1cCare and the Variant II Turbo (Bio-Rad Laboratories, Inc.; Hercules, CA, USA) in 14 days.
The SD A1cCare POC HbA1c analyzer showed excellent precision, linearity and correlation with Variant II Turbo high-performance liquid chromatography (HPLC) and it showed excellent accuracy with IFCC targets. Deviations from IFCC targets at 30, 60, and 90 mmol/mol IFCC levels were − 1.95, − 1.85, and − 1.74 mmol/mol, respectively. The coefficients of variation based on EP9-A2 protocol were 2.6% in SI units and 1.8% in NGSP units.
The authors concluded that the SD A1cCare analyzer showed excellent precision, linearity, correlation with the Variant II Turbo analyzer, and accuracy with IFCC targets and therefore, it may be suitable for HbA1c assays in the POC setting and in small laboratories. The study was published in the June, 2015 issue of the journal Clinical Biochemistry.
Related Links:
Seoul National University College of Medicine
SD Biosensor Inc.
Bio-Rad Laboratories
The immediate feedback of HbA1c levels is highly effective for controlling plasma glucose levels and point-of-care (POC) instruments, characterized as fast, portable, and easy-to-use, have been shown to be suitable for providing rapid feedback of HbA1c levels.
Medical laboratory scientists at the Seoul National University College of Medicine (Republic of Korea) assessed the analytical performance of a newly developed POC HbA1c analyzer. To investigate the accuracy of the analyzer, they measured the deviation from the International Federation of Clinical Chemistry (IFCC) reference targets using two sets of 12 duplicate fresh venous whole blood specimens with a range of 33.3 mmol/mol to 91.3 mmol/mol in International System of Units (SI) units, or 5.2% to 10.5% in National Glycohemoglobin Standardization Program (NSGP) units.
The SD A1cCare analyzer (SD Biosensor, Inc.; Suwon, Korea), which utilizes an immunochromatography assay, was used to measures total hemoglobin and HbA1c levels via its optical system and calculates the percentage HbA1c. The SD A1cCare Test kit uses an anti-HbA1c antibody that is specific for the first few amino acid residues of the glycated N-terminus of the β-chain of hemoglobin A0. The scientists tested 150 residual specimens using both the SD A1cCare and the Variant II Turbo (Bio-Rad Laboratories, Inc.; Hercules, CA, USA) in 14 days.
The SD A1cCare POC HbA1c analyzer showed excellent precision, linearity and correlation with Variant II Turbo high-performance liquid chromatography (HPLC) and it showed excellent accuracy with IFCC targets. Deviations from IFCC targets at 30, 60, and 90 mmol/mol IFCC levels were − 1.95, − 1.85, and − 1.74 mmol/mol, respectively. The coefficients of variation based on EP9-A2 protocol were 2.6% in SI units and 1.8% in NGSP units.
The authors concluded that the SD A1cCare analyzer showed excellent precision, linearity, correlation with the Variant II Turbo analyzer, and accuracy with IFCC targets and therefore, it may be suitable for HbA1c assays in the POC setting and in small laboratories. The study was published in the June, 2015 issue of the journal Clinical Biochemistry.
Related Links:
Seoul National University College of Medicine
SD Biosensor Inc.
Bio-Rad Laboratories
Read the full article by registering today, it's FREE!
Register now for FREE to LabMedica.com and get access to news and events that shape the world of Clinical Laboratory Medicine.
- Free digital version edition of LabMedica International sent by email on regular basis
- Free print version of LabMedica International magazine (available only outside USA and Canada).
- Free and unlimited access to back issues of LabMedica International in digital format
- Free LabMedica International Newsletter sent every week containing the latest news
- Free breaking news sent via email
- Free access to Events Calendar
- Free access to LinkXpress new product services
- REGISTRATION IS FREE AND EASY!
Sign in: Registered website members
Sign in: Registered magazine subscribers
Latest Clinical Chem. News
- New PSA-Based Prognostic Model Improves Prostate Cancer Risk Assessment
- Extracellular Vesicles Linked to Heart Failure Risk in CKD Patients
- Study Compares Analytical Performance of Quantitative Hepatitis B Surface Antigen Assays
- Blood Test Could Predict and Identify Early Relapses in Myeloma Patients
- Compact Raman Imaging System Detects Subtle Tumor Signals
- Noninvasive Blood-Glucose Monitoring to Replace Finger Pricks for Diabetics
- POC Breath Diagnostic System to Detect Pneumonia-Causing Pathogens
- Online Tool Detects Drug Exposure Directly from Patient Samples
- Chemical Imaging Probe Could Track and Treat Prostate Cancer
- Mismatch Between Two Common Kidney Function Tests Indicates Serious Health Problems
- VOCs Show Promise for Early Multi-Cancer Detection
- Portable Raman Spectroscopy Offers Cost-Effective Kidney Disease Diagnosis at POC
- Gold Nanoparticles to Improve Accuracy of Ovarian Cancer Diagnosis
- Simultaneous Cell Isolation Technology Improves Cancer Diagnostic Accuracy
- Simple Non-Invasive Hair-Based Test Could Speed ALS Diagnosis
- Paper Strip Saliva Test Detects Elevated Uric Acid Levels Without Blood Draws
Channels
Molecular Diagnostics
view channel
Diagnostic Device Predicts Treatment Response for Brain Tumors Via Blood Test
Glioblastoma is one of the deadliest forms of brain cancer, largely because doctors have no reliable way to determine whether treatments are working in real time. Assessing therapeutic response currently... Read more
Blood Test Detects Early-Stage Cancers by Measuring Epigenetic Instability
Early-stage cancers are notoriously difficult to detect because molecular changes are subtle and often missed by existing screening tools. Many liquid biopsies rely on measuring absolute DNA methylation... Read more
“Lab-On-A-Disc” Device Paves Way for More Automated Liquid Biopsies
Extracellular vesicles (EVs) are tiny particles released by cells into the bloodstream that carry molecular information about a cell’s condition, including whether it is cancerous. However, EVs are highly... Read more
Blood Test Identifies Inflammatory Breast Cancer Patients at Increased Risk of Brain Metastasis
Brain metastasis is a frequent and devastating complication in patients with inflammatory breast cancer, an aggressive subtype with limited treatment options. Despite its high incidence, the biological... Read moreHematology
view channel
New Guidelines Aim to Improve AL Amyloidosis Diagnosis
Light chain (AL) amyloidosis is a rare, life-threatening bone marrow disorder in which abnormal amyloid proteins accumulate in organs. Approximately 3,260 people in the United States are diagnosed... Read more
Fast and Easy Test Could Revolutionize Blood Transfusions
Blood transfusions are a cornerstone of modern medicine, yet red blood cells can deteriorate quietly while sitting in cold storage for weeks. Although blood units have a fixed expiration date, cells from... Read more
Automated Hemostasis System Helps Labs of All Sizes Optimize Workflow
High-volume hemostasis sections must sustain rapid turnaround while managing reruns and reflex testing. Manual tube handling and preanalytical checks can strain staff time and increase opportunities for error.... Read more
High-Sensitivity Blood Test Improves Assessment of Clotting Risk in Heart Disease Patients
Blood clotting is essential for preventing bleeding, but even small imbalances can lead to serious conditions such as thrombosis or dangerous hemorrhage. In cardiovascular disease, clinicians often struggle... Read moreImmunology
view channelBlood Test Identifies Lung Cancer Patients Who Can Benefit from Immunotherapy Drug
Small cell lung cancer (SCLC) is an aggressive disease with limited treatment options, and even newly approved immunotherapies do not benefit all patients. While immunotherapy can extend survival for some,... Read more
Whole-Genome Sequencing Approach Identifies Cancer Patients Benefitting From PARP-Inhibitor Treatment
Targeted cancer therapies such as PARP inhibitors can be highly effective, but only for patients whose tumors carry specific DNA repair defects. Identifying these patients accurately remains challenging,... Read more
Ultrasensitive Liquid Biopsy Demonstrates Efficacy in Predicting Immunotherapy Response
Immunotherapy has transformed cancer treatment, but only a small proportion of patients experience lasting benefit, with response rates often remaining between 10% and 20%. Clinicians currently lack reliable... Read moreMicrobiology
view channel
Comprehensive Review Identifies Gut Microbiome Signatures Associated With Alzheimer’s Disease
Alzheimer’s disease affects approximately 6.7 million people in the United States and nearly 50 million worldwide, yet early cognitive decline remains difficult to characterize. Increasing evidence suggests... Read moreAI-Powered Platform Enables Rapid Detection of Drug-Resistant C. Auris Pathogens
Infections caused by the pathogenic yeast Candida auris pose a significant threat to hospitalized patients, particularly those with weakened immune systems or those who have invasive medical devices.... Read morePathology
view channel
Engineered Yeast Cells Enable Rapid Testing of Cancer Immunotherapy
Developing new cancer immunotherapies is a slow, costly, and high-risk process, particularly for CAR T cell treatments that must precisely recognize cancer-specific antigens. Small differences in tumor... Read more
First-Of-Its-Kind Test Identifies Autism Risk at Birth
Autism spectrum disorder is treatable, and extensive research shows that early intervention can significantly improve cognitive, social, and behavioral outcomes. Yet in the United States, the average age... Read moreTechnology
view channel
Robotic Technology Unveiled for Automated Diagnostic Blood Draws
Routine diagnostic blood collection is a high‑volume task that can strain staffing and introduce human‑dependent variability, with downstream implications for sample quality and patient experience.... Read more
ADLM Launches First-of-Its-Kind Data Science Program for Laboratory Medicine Professionals
Clinical laboratories generate billions of test results each year, creating a treasure trove of data with the potential to support more personalized testing, improve operational efficiency, and enhance patient care.... Read moreAptamer Biosensor Technology to Transform Virus Detection
Rapid and reliable virus detection is essential for controlling outbreaks, from seasonal influenza to global pandemics such as COVID-19. Conventional diagnostic methods, including cell culture, antigen... Read more
AI Models Could Predict Pre-Eclampsia and Anemia Earlier Using Routine Blood Tests
Pre-eclampsia and anemia are major contributors to maternal and child mortality worldwide, together accounting for more than half a million deaths each year and leaving millions with long-term health complications.... Read moreIndustry
view channelNew Collaboration Brings Automated Mass Spectrometry to Routine Laboratory Testing
Mass spectrometry is a powerful analytical technique that identifies and quantifies molecules based on their mass and electrical charge. Its high selectivity, sensitivity, and accuracy make it indispensable... Read more
AI-Powered Cervical Cancer Test Set for Major Rollout in Latin America
Noul Co., a Korean company specializing in AI-based blood and cancer diagnostics, announced it will supply its intelligence (AI)-based miLab CER cervical cancer diagnostic solution to Mexico under a multi‑year... Read more
Diasorin and Fisher Scientific Enter into US Distribution Agreement for Molecular POC Platform
Diasorin (Saluggia, Italy) has entered into an exclusive distribution agreement with Fisher Scientific, part of Thermo Fisher Scientific (Waltham, MA, USA), for the LIAISON NES molecular point-of-care... Read more








 Analyzer.jpg)