Performance of Five Lipoprotein(a) Immunoassays Evaluated
|
By LabMedica International staff writers Posted on 06 Apr 2021 |

Image: The Diazyme Lipoprotein A assay kit is for the in vitro quantitative determination of Lp(a) concentration in human serum or plasma on Clinical Chemistry Systems (Photo courtesy of Diazyme)
Increased serum Lipoprotein(a) [Lp(a)] is considered an independent, inherited risk factor for atherosclerotic cardiovascular disease (ASCVD), including myocardial infarction and stroke. This risk is believed to be due to pro-atherogenic, pro-inflammatory, and pro-thrombotic properties of the Lp(a) particle.
Lp(a) can be measured using a variety of methods, including enzyme linked immunosorbent assays (ELISA), latex-agglutination, immunoturbidimetry/immunonephelometry, electrophoresis, and immunofixation electrophoresis (IFE). A confounding factor is that Lp(a) can be reported in either mass units (mg/dL of the total Lp(a) particle) or molar concentration of particles (nmol/L).
Medical Laboratorians at the ARUP Institute of Clinical and Experimental Pathology (Salt Lake City, UT, USA) and their colleagues used 80 residual serum samples, and specimens from 120 self-reported healthy subjects (61 females/59 males) were also tested in a study. The aim of the study was to evaluate performance characteristic of five different Lp(a) assays using the cobas c501 analyzer (Roche Diagnostics, Indianapolis, IN, USA).
The five assay for Lp(a) used were from: Diazyme (Poway, CA, USA), Kamiya Biomedical Company (Seattle, WA, USA), MedTest Dx (Canton, MI, USA), Randox Laboratories (Crumlin, UK) and Roche Diagnostics, configured to mg/dL units. Assays from Diazyme and Kamiya were also configured using nmol/L units in separate studies. Studies included sensitivity, imprecision, linearity, method comparison, and evaluation of healthy subjects. Imprecision (intra-day, 20 replicates; inter-day, duplicates twice daily for five days) and linearity were evaluated using patient pools. Linearity assessed a minimum of five patient splits spanning the analytical measurement range (AMR).
The scientists reported that all methods met manufacturer claims regarding sensitivity: observed (manufacturer claim): Diazyme, 0.7 mg/dL (1.3 mg/dL); Kamiya, 1.2 mg/dL (5.0 mg/dL); MedTest, 0.2 mg/dL (1.3 mg/dL); Randox, 0.7 mg/dL (3.0 mg/dL); Roche, 0.3 mg/dL (4.0 mg/dL). Lp(a) assays also demonstrated acceptable imprecision and met manufacturers’ claims, with CVs less than 6% in all cases. Imprecision studies demonstrated %CVs ranging from 2.5 – 5.2% for the low pool (average concentration from 7.3 – 12.4 mg/dL); high pool %CVs ranged from 0.8 – 3.0% (average concentrations from (31.5 – 50.2 mg/dL). Linearity was confirmed for all assays. Variation in accuracy was observed when comparing results to an all method average. Lp(s) results were higher in females versus males in self-reported healthy subjects.
The authors concluded that all assays performed according to manufacturer described performance characteristics, although differences were observed across Lp(a) assays tested when compared to an all method average. The study was published on March 24, 2021 in the journal Practical Laboratory Medicine.
Related Links:
ARUP Institute of Clinical and Experimental Pathology
Roche Diagnostics
Diazyme
Kamiya Biomedical Company
MedTest Dx
Randox Laboratories
Lp(a) can be measured using a variety of methods, including enzyme linked immunosorbent assays (ELISA), latex-agglutination, immunoturbidimetry/immunonephelometry, electrophoresis, and immunofixation electrophoresis (IFE). A confounding factor is that Lp(a) can be reported in either mass units (mg/dL of the total Lp(a) particle) or molar concentration of particles (nmol/L).
Medical Laboratorians at the ARUP Institute of Clinical and Experimental Pathology (Salt Lake City, UT, USA) and their colleagues used 80 residual serum samples, and specimens from 120 self-reported healthy subjects (61 females/59 males) were also tested in a study. The aim of the study was to evaluate performance characteristic of five different Lp(a) assays using the cobas c501 analyzer (Roche Diagnostics, Indianapolis, IN, USA).
The five assay for Lp(a) used were from: Diazyme (Poway, CA, USA), Kamiya Biomedical Company (Seattle, WA, USA), MedTest Dx (Canton, MI, USA), Randox Laboratories (Crumlin, UK) and Roche Diagnostics, configured to mg/dL units. Assays from Diazyme and Kamiya were also configured using nmol/L units in separate studies. Studies included sensitivity, imprecision, linearity, method comparison, and evaluation of healthy subjects. Imprecision (intra-day, 20 replicates; inter-day, duplicates twice daily for five days) and linearity were evaluated using patient pools. Linearity assessed a minimum of five patient splits spanning the analytical measurement range (AMR).
The scientists reported that all methods met manufacturer claims regarding sensitivity: observed (manufacturer claim): Diazyme, 0.7 mg/dL (1.3 mg/dL); Kamiya, 1.2 mg/dL (5.0 mg/dL); MedTest, 0.2 mg/dL (1.3 mg/dL); Randox, 0.7 mg/dL (3.0 mg/dL); Roche, 0.3 mg/dL (4.0 mg/dL). Lp(a) assays also demonstrated acceptable imprecision and met manufacturers’ claims, with CVs less than 6% in all cases. Imprecision studies demonstrated %CVs ranging from 2.5 – 5.2% for the low pool (average concentration from 7.3 – 12.4 mg/dL); high pool %CVs ranged from 0.8 – 3.0% (average concentrations from (31.5 – 50.2 mg/dL). Linearity was confirmed for all assays. Variation in accuracy was observed when comparing results to an all method average. Lp(s) results were higher in females versus males in self-reported healthy subjects.
The authors concluded that all assays performed according to manufacturer described performance characteristics, although differences were observed across Lp(a) assays tested when compared to an all method average. The study was published on March 24, 2021 in the journal Practical Laboratory Medicine.
Related Links:
ARUP Institute of Clinical and Experimental Pathology
Roche Diagnostics
Diazyme
Kamiya Biomedical Company
MedTest Dx
Randox Laboratories
Latest Clinical Chem. News
- New PSA-Based Prognostic Model Improves Prostate Cancer Risk Assessment
- Extracellular Vesicles Linked to Heart Failure Risk in CKD Patients
- Study Compares Analytical Performance of Quantitative Hepatitis B Surface Antigen Assays
- Blood Test Could Predict and Identify Early Relapses in Myeloma Patients
- Compact Raman Imaging System Detects Subtle Tumor Signals
- Noninvasive Blood-Glucose Monitoring to Replace Finger Pricks for Diabetics
- POC Breath Diagnostic System to Detect Pneumonia-Causing Pathogens
- Online Tool Detects Drug Exposure Directly from Patient Samples
- Chemical Imaging Probe Could Track and Treat Prostate Cancer
- Mismatch Between Two Common Kidney Function Tests Indicates Serious Health Problems
- VOCs Show Promise for Early Multi-Cancer Detection
- Portable Raman Spectroscopy Offers Cost-Effective Kidney Disease Diagnosis at POC
- Gold Nanoparticles to Improve Accuracy of Ovarian Cancer Diagnosis
- Simultaneous Cell Isolation Technology Improves Cancer Diagnostic Accuracy
- Simple Non-Invasive Hair-Based Test Could Speed ALS Diagnosis
- Paper Strip Saliva Test Detects Elevated Uric Acid Levels Without Blood Draws
Channels
Clinical Chemistry
view channel
New PSA-Based Prognostic Model Improves Prostate Cancer Risk Assessment
Prostate cancer is the second-leading cause of cancer death among American men, and about one in eight will be diagnosed in their lifetime. Screening relies on blood levels of prostate-specific antigen... Read more
Extracellular Vesicles Linked to Heart Failure Risk in CKD Patients
Chronic kidney disease (CKD) affects more than 1 in 7 Americans and is strongly associated with cardiovascular complications, which account for more than half of deaths among people with CKD.... Read moreMolecular Diagnostics
view channel
Diagnostic Device Predicts Treatment Response for Brain Tumors Via Blood Test
Glioblastoma is one of the deadliest forms of brain cancer, largely because doctors have no reliable way to determine whether treatments are working in real time. Assessing therapeutic response currently... Read more
Blood Test Detects Early-Stage Cancers by Measuring Epigenetic Instability
Early-stage cancers are notoriously difficult to detect because molecular changes are subtle and often missed by existing screening tools. Many liquid biopsies rely on measuring absolute DNA methylation... Read more
“Lab-On-A-Disc” Device Paves Way for More Automated Liquid Biopsies
Extracellular vesicles (EVs) are tiny particles released by cells into the bloodstream that carry molecular information about a cell’s condition, including whether it is cancerous. However, EVs are highly... Read more
Blood Test Identifies Inflammatory Breast Cancer Patients at Increased Risk of Brain Metastasis
Brain metastasis is a frequent and devastating complication in patients with inflammatory breast cancer, an aggressive subtype with limited treatment options. Despite its high incidence, the biological... Read moreHematology
view channel
New Guidelines Aim to Improve AL Amyloidosis Diagnosis
Light chain (AL) amyloidosis is a rare, life-threatening bone marrow disorder in which abnormal amyloid proteins accumulate in organs. Approximately 3,260 people in the United States are diagnosed... Read more
Fast and Easy Test Could Revolutionize Blood Transfusions
Blood transfusions are a cornerstone of modern medicine, yet red blood cells can deteriorate quietly while sitting in cold storage for weeks. Although blood units have a fixed expiration date, cells from... Read more
Automated Hemostasis System Helps Labs of All Sizes Optimize Workflow
High-volume hemostasis sections must sustain rapid turnaround while managing reruns and reflex testing. Manual tube handling and preanalytical checks can strain staff time and increase opportunities for error.... Read more
High-Sensitivity Blood Test Improves Assessment of Clotting Risk in Heart Disease Patients
Blood clotting is essential for preventing bleeding, but even small imbalances can lead to serious conditions such as thrombosis or dangerous hemorrhage. In cardiovascular disease, clinicians often struggle... Read moreMicrobiology
view channel
Comprehensive Review Identifies Gut Microbiome Signatures Associated With Alzheimer’s Disease
Alzheimer’s disease affects approximately 6.7 million people in the United States and nearly 50 million worldwide, yet early cognitive decline remains difficult to characterize. Increasing evidence suggests... Read moreAI-Powered Platform Enables Rapid Detection of Drug-Resistant C. Auris Pathogens
Infections caused by the pathogenic yeast Candida auris pose a significant threat to hospitalized patients, particularly those with weakened immune systems or those who have invasive medical devices.... Read morePathology
view channel
Engineered Yeast Cells Enable Rapid Testing of Cancer Immunotherapy
Developing new cancer immunotherapies is a slow, costly, and high-risk process, particularly for CAR T cell treatments that must precisely recognize cancer-specific antigens. Small differences in tumor... Read more
First-Of-Its-Kind Test Identifies Autism Risk at Birth
Autism spectrum disorder is treatable, and extensive research shows that early intervention can significantly improve cognitive, social, and behavioral outcomes. Yet in the United States, the average age... Read moreTechnology
view channel
Robotic Technology Unveiled for Automated Diagnostic Blood Draws
Routine diagnostic blood collection is a high‑volume task that can strain staffing and introduce human‑dependent variability, with downstream implications for sample quality and patient experience.... Read more
ADLM Launches First-of-Its-Kind Data Science Program for Laboratory Medicine Professionals
Clinical laboratories generate billions of test results each year, creating a treasure trove of data with the potential to support more personalized testing, improve operational efficiency, and enhance patient care.... Read moreAptamer Biosensor Technology to Transform Virus Detection
Rapid and reliable virus detection is essential for controlling outbreaks, from seasonal influenza to global pandemics such as COVID-19. Conventional diagnostic methods, including cell culture, antigen... Read more
AI Models Could Predict Pre-Eclampsia and Anemia Earlier Using Routine Blood Tests
Pre-eclampsia and anemia are major contributors to maternal and child mortality worldwide, together accounting for more than half a million deaths each year and leaving millions with long-term health complications.... Read moreIndustry
view channelNew Collaboration Brings Automated Mass Spectrometry to Routine Laboratory Testing
Mass spectrometry is a powerful analytical technique that identifies and quantifies molecules based on their mass and electrical charge. Its high selectivity, sensitivity, and accuracy make it indispensable... Read more
AI-Powered Cervical Cancer Test Set for Major Rollout in Latin America
Noul Co., a Korean company specializing in AI-based blood and cancer diagnostics, announced it will supply its intelligence (AI)-based miLab CER cervical cancer diagnostic solution to Mexico under a multi‑year... Read more
Diasorin and Fisher Scientific Enter into US Distribution Agreement for Molecular POC Platform
Diasorin (Saluggia, Italy) has entered into an exclusive distribution agreement with Fisher Scientific, part of Thermo Fisher Scientific (Waltham, MA, USA), for the LIAISON NES molecular point-of-care... Read more