Performance of Five Lipoprotein(a) Immunoassays Evaluated
|
By LabMedica International staff writers Posted on 06 Apr 2021 |

Image: The Diazyme Lipoprotein A assay kit is for the in vitro quantitative determination of Lp(a) concentration in human serum or plasma on Clinical Chemistry Systems (Photo courtesy of Diazyme)
Increased serum Lipoprotein(a) [Lp(a)] is considered an independent, inherited risk factor for atherosclerotic cardiovascular disease (ASCVD), including myocardial infarction and stroke. This risk is believed to be due to pro-atherogenic, pro-inflammatory, and pro-thrombotic properties of the Lp(a) particle.
Lp(a) can be measured using a variety of methods, including enzyme linked immunosorbent assays (ELISA), latex-agglutination, immunoturbidimetry/immunonephelometry, electrophoresis, and immunofixation electrophoresis (IFE). A confounding factor is that Lp(a) can be reported in either mass units (mg/dL of the total Lp(a) particle) or molar concentration of particles (nmol/L).
Medical Laboratorians at the ARUP Institute of Clinical and Experimental Pathology (Salt Lake City, UT, USA) and their colleagues used 80 residual serum samples, and specimens from 120 self-reported healthy subjects (61 females/59 males) were also tested in a study. The aim of the study was to evaluate performance characteristic of five different Lp(a) assays using the cobas c501 analyzer (Roche Diagnostics, Indianapolis, IN, USA).
The five assay for Lp(a) used were from: Diazyme (Poway, CA, USA), Kamiya Biomedical Company (Seattle, WA, USA), MedTest Dx (Canton, MI, USA), Randox Laboratories (Crumlin, UK) and Roche Diagnostics, configured to mg/dL units. Assays from Diazyme and Kamiya were also configured using nmol/L units in separate studies. Studies included sensitivity, imprecision, linearity, method comparison, and evaluation of healthy subjects. Imprecision (intra-day, 20 replicates; inter-day, duplicates twice daily for five days) and linearity were evaluated using patient pools. Linearity assessed a minimum of five patient splits spanning the analytical measurement range (AMR).
The scientists reported that all methods met manufacturer claims regarding sensitivity: observed (manufacturer claim): Diazyme, 0.7 mg/dL (1.3 mg/dL); Kamiya, 1.2 mg/dL (5.0 mg/dL); MedTest, 0.2 mg/dL (1.3 mg/dL); Randox, 0.7 mg/dL (3.0 mg/dL); Roche, 0.3 mg/dL (4.0 mg/dL). Lp(a) assays also demonstrated acceptable imprecision and met manufacturers’ claims, with CVs less than 6% in all cases. Imprecision studies demonstrated %CVs ranging from 2.5 – 5.2% for the low pool (average concentration from 7.3 – 12.4 mg/dL); high pool %CVs ranged from 0.8 – 3.0% (average concentrations from (31.5 – 50.2 mg/dL). Linearity was confirmed for all assays. Variation in accuracy was observed when comparing results to an all method average. Lp(s) results were higher in females versus males in self-reported healthy subjects.
The authors concluded that all assays performed according to manufacturer described performance characteristics, although differences were observed across Lp(a) assays tested when compared to an all method average. The study was published on March 24, 2021 in the journal Practical Laboratory Medicine.
Related Links:
ARUP Institute of Clinical and Experimental Pathology
Roche Diagnostics
Diazyme
Kamiya Biomedical Company
MedTest Dx
Randox Laboratories
Lp(a) can be measured using a variety of methods, including enzyme linked immunosorbent assays (ELISA), latex-agglutination, immunoturbidimetry/immunonephelometry, electrophoresis, and immunofixation electrophoresis (IFE). A confounding factor is that Lp(a) can be reported in either mass units (mg/dL of the total Lp(a) particle) or molar concentration of particles (nmol/L).
Medical Laboratorians at the ARUP Institute of Clinical and Experimental Pathology (Salt Lake City, UT, USA) and their colleagues used 80 residual serum samples, and specimens from 120 self-reported healthy subjects (61 females/59 males) were also tested in a study. The aim of the study was to evaluate performance characteristic of five different Lp(a) assays using the cobas c501 analyzer (Roche Diagnostics, Indianapolis, IN, USA).
The five assay for Lp(a) used were from: Diazyme (Poway, CA, USA), Kamiya Biomedical Company (Seattle, WA, USA), MedTest Dx (Canton, MI, USA), Randox Laboratories (Crumlin, UK) and Roche Diagnostics, configured to mg/dL units. Assays from Diazyme and Kamiya were also configured using nmol/L units in separate studies. Studies included sensitivity, imprecision, linearity, method comparison, and evaluation of healthy subjects. Imprecision (intra-day, 20 replicates; inter-day, duplicates twice daily for five days) and linearity were evaluated using patient pools. Linearity assessed a minimum of five patient splits spanning the analytical measurement range (AMR).
The scientists reported that all methods met manufacturer claims regarding sensitivity: observed (manufacturer claim): Diazyme, 0.7 mg/dL (1.3 mg/dL); Kamiya, 1.2 mg/dL (5.0 mg/dL); MedTest, 0.2 mg/dL (1.3 mg/dL); Randox, 0.7 mg/dL (3.0 mg/dL); Roche, 0.3 mg/dL (4.0 mg/dL). Lp(a) assays also demonstrated acceptable imprecision and met manufacturers’ claims, with CVs less than 6% in all cases. Imprecision studies demonstrated %CVs ranging from 2.5 – 5.2% for the low pool (average concentration from 7.3 – 12.4 mg/dL); high pool %CVs ranged from 0.8 – 3.0% (average concentrations from (31.5 – 50.2 mg/dL). Linearity was confirmed for all assays. Variation in accuracy was observed when comparing results to an all method average. Lp(s) results were higher in females versus males in self-reported healthy subjects.
The authors concluded that all assays performed according to manufacturer described performance characteristics, although differences were observed across Lp(a) assays tested when compared to an all method average. The study was published on March 24, 2021 in the journal Practical Laboratory Medicine.
Related Links:
ARUP Institute of Clinical and Experimental Pathology
Roche Diagnostics
Diazyme
Kamiya Biomedical Company
MedTest Dx
Randox Laboratories
Latest Immunology News
- New Test Distinguishes Vaccine-Induced False Positives from Active HIV Infection
- Gene Signature Test Predicts Response to Key Breast Cancer Treatment
- Chip Captures Cancer Cells from Blood to Help Select Right Breast Cancer Treatment
- Blood-Based Liquid Biopsy Model Analyzes Immunotherapy Effectiveness
- Signature Genes Predict T-Cell Expansion in Cancer Immunotherapy
- Molecular Microscope Diagnostic System Assesses Lung Transplant Rejection
- Blood Test Tracks Treatment Resistance in High-Grade Serous Ovarian Cancer
- Luminescent Probe Measures Immune Cell Activity in Real Time
- Blood-Based Immune Cell Signatures Could Guide Treatment Decisions for Critically Ill Patients
- Novel Tool Predicts Most Effective Multiple Sclerosis Medication for Patients
- Companion Diagnostic Test for CRC Patients Identifies Eligible Treatment Population
- Novel Tool Uses Deep Learning for Precision Cancer Therapy
- Companion Diagnostic Test Identifies HER2-Ultralow Breast Cancer and Biliary Tract Cancer Patients
- Novel Multiplex Assay Supports Diagnosis of Autoimmune Vasculitis
- Blood Test Predicts Immunotherapy Efficacy in Triple-Negative Breast Cancer
- Simple Genetic Testing Could Predict Treatment Success in Multiple Sclerosis Patients
Channels
Clinical Chemistry
view channel
Chemical Imaging Probe Could Track and Treat Prostate Cancer
Prostate cancer remains a leading cause of illness and death among men, with many patients eventually developing resistance to standard hormone-blocking therapies. These drugs often lose effectiveness... Read more
Mismatch Between Two Common Kidney Function Tests Indicates Serious Health Problems
Creatinine has long been the standard for measuring kidney filtration, while cystatin C — a protein produced by all human cells — has been recommended as a complementary marker because it is influenced... Read moreMolecular Diagnostics
view channel
Simple Urine Test to Revolutionize Bladder Cancer Diagnosis and Treatment
Bladder cancer is one of the most common and deadly urological cancers and is marked by a high rate of recurrence. Diagnosis and follow-up still rely heavily on invasive cystoscopy or urine cytology, which... Read more
Blood Test to Enable Earlier and Simpler Detection of Liver Fibrosis
Persistent liver damage caused by alcohol misuse or viral infections can trigger liver fibrosis, a condition in which healthy tissue is gradually replaced by collagen fibers. Even after successful treatment... Read moreHematology
view channel
Platelet Activity Blood Test in Middle Age Could Identify Early Alzheimer’s Risk
Early detection of Alzheimer’s disease remains one of the biggest unmet needs in neurology, particularly because the biological changes underlying the disorder begin decades before memory symptoms appear.... Read more
Microvesicles Measurement Could Detect Vascular Injury in Sickle Cell Disease Patients
Assessing disease severity in sickle cell disease (SCD) remains challenging, especially when trying to predict hemolysis, vascular injury, and risk of complications such as vaso-occlusive crises.... Read more
ADLM’s New Coagulation Testing Guidance to Improve Care for Patients on Blood Thinners
Direct oral anticoagulants (DOACs) are one of the most common types of blood thinners. Patients take them to prevent a host of complications that could arise from blood clotting, including stroke, deep... Read moreMicrobiology
view channel
Rapid Diagnostic Test Matches Gold Standard for Sepsis Detection
Sepsis kills 11 million people worldwide every year and generates massive healthcare costs. In the USA and Europe alone, sepsis accounts for USD 100 billion in annual hospitalization expenses.... Read moreRapid POC Tuberculosis Test Provides Results Within 15 Minutes
Tuberculosis remains one of the world’s deadliest infectious diseases, and reducing new cases depends on identifying individuals with latent infection before it progresses. Current diagnostic tools often... Read more
Rapid Assay Identifies Bloodstream Infection Pathogens Directly from Patient Samples
Bloodstream infections in sepsis progress quickly and demand rapid, precise diagnosis. Current blood-culture methods often take one to five days to identify the pathogen, leaving clinicians to treat blindly... Read morePathology
view channel
Tunable Cell-Sorting Device Holds Potential for Multiple Biomedical Applications
Isolating rare cancer cells from blood is essential for diagnosing metastasis and guiding treatment decisions, but remains technically challenging. Many existing techniques struggle to balance accuracy,... Read moreAI Tool Outperforms Doctors in Spotting Blood Cell Abnormalities
Diagnosing blood disorders depends on recognizing subtle abnormalities in cell size, shape, and structure, yet this process is slow, subjective, and requires years of expert training. Even specialists... Read moreTechnology
view channel
Artificial Intelligence Model Could Accelerate Rare Disease Diagnosis
Identifying which genetic variants actually cause disease remains one of the biggest challenges in genomic medicine. Each person carries tens of thousands of DNA changes, yet only a few meaningfully alter... Read more
AI Saliva Sensor Enables Early Detection of Head and Neck Cancer
Early detection of head and neck cancer remains difficult because the disease produces few or no symptoms in its earliest stages, and lesions often lie deep within the head or neck, where biopsy or endoscopy... Read moreIndustry
view channel
Abbott Acquires Cancer-Screening Company Exact Sciences
Abbott (Abbott Park, IL, USA) has entered into a definitive agreement to acquire Exact Sciences (Madison, WI, USA), enabling it to enter and lead in fast-growing cancer diagnostics segments.... Read more