Performance of Immunosuppressant Drug Assay Kits Analyzed
By LabMedica International staff writers Posted on 08 Aug 2017 |

Image: The cobas e 411immunoassay analyzer (Photo courtesy of Roche Diagnostics).
Cyclosporine (CsA) and tacrolimus (TAC) are immunosuppressant drugs that are often used to treat autoimmune diseases and as transplantation therapy and therefore, their concentrations need to be monitored carefully.
The main analytical methods for CsA and TAC in whole blood have been liquid chromatography combined with mass spectrometric detection (LC-MS/MS), affinity chrome-mediated immunoassay (ACMIA), and chemiluminescence immunoassays (CLIA).
Scientists at the Osaka University Hospital (Osaka, Japan) used 200 residual EDTA whole blood samples from patients who had received an organ transplant and were under CsA or TAC therapy. CsA concentrations were measured using an affinity chrome-mediated immunoassay (ACMIA) and an electrochemiluminescence immunoassay (ECLIA). TAC concentrations were measured using a chemiluminescence immunoassay (CLIA) and ECLIA.
The investigators used the Elecsys Cyclosporine assay kit and the Elecsys Tacrolimus assay kit where the concentrations were measured on a cobas e411 analyzer. The ACMIA for measurement of CsA concentration was performed with a Dimension Xpand analyzer and CLIA for measurement of TAC concentration with an ARCHITECT i2000SR. It was necessary for accurate measurements of CsA and TAC concentrations to sufficiently mix samples and pretreatment reagent. The hematocrit was assayed using the fully automated hematological analyzer Sysmex XN-1000.
The scientists investigated assay precision, linearity, lower limit of quantitation (LOQ), stability of calibration, influence of interference substances and the hematocrit, correlation of ACMIA with ECLIA, and correlation of CLIA with ECLIA. They found that the limits of quantitation (LOQ) were defined as the concentration at which the coefficient of variation (CV) was approximately 10%. Each lower LOQ obtained was 16 ng/mL (CsA), and 0.95 ng/mL (TAC). CsA and TAC calibrations were stable for at least 21 days. Neither the presence of conjugated bilirubin, unconjugated bilirubin, chyle, and rheumatoid factor nor the hematocrit affected these assays.
The authors concluded that the analytical performances of the Elecsys Cyclosporine and Elecsys Tacrolimus assays were acceptable. Furthermore, CsA and TAC concentrations may be simultaneously measured using a single pretreatment, which is of benefit if patients have to undertake conversion between these two drugs. Additionally, it benefits the workflow in the clinical laboratory. The study was published in the August 2017 issue of the journal Practical Laboratory Medicine.
The main analytical methods for CsA and TAC in whole blood have been liquid chromatography combined with mass spectrometric detection (LC-MS/MS), affinity chrome-mediated immunoassay (ACMIA), and chemiluminescence immunoassays (CLIA).
Scientists at the Osaka University Hospital (Osaka, Japan) used 200 residual EDTA whole blood samples from patients who had received an organ transplant and were under CsA or TAC therapy. CsA concentrations were measured using an affinity chrome-mediated immunoassay (ACMIA) and an electrochemiluminescence immunoassay (ECLIA). TAC concentrations were measured using a chemiluminescence immunoassay (CLIA) and ECLIA.
The investigators used the Elecsys Cyclosporine assay kit and the Elecsys Tacrolimus assay kit where the concentrations were measured on a cobas e411 analyzer. The ACMIA for measurement of CsA concentration was performed with a Dimension Xpand analyzer and CLIA for measurement of TAC concentration with an ARCHITECT i2000SR. It was necessary for accurate measurements of CsA and TAC concentrations to sufficiently mix samples and pretreatment reagent. The hematocrit was assayed using the fully automated hematological analyzer Sysmex XN-1000.
The scientists investigated assay precision, linearity, lower limit of quantitation (LOQ), stability of calibration, influence of interference substances and the hematocrit, correlation of ACMIA with ECLIA, and correlation of CLIA with ECLIA. They found that the limits of quantitation (LOQ) were defined as the concentration at which the coefficient of variation (CV) was approximately 10%. Each lower LOQ obtained was 16 ng/mL (CsA), and 0.95 ng/mL (TAC). CsA and TAC calibrations were stable for at least 21 days. Neither the presence of conjugated bilirubin, unconjugated bilirubin, chyle, and rheumatoid factor nor the hematocrit affected these assays.
The authors concluded that the analytical performances of the Elecsys Cyclosporine and Elecsys Tacrolimus assays were acceptable. Furthermore, CsA and TAC concentrations may be simultaneously measured using a single pretreatment, which is of benefit if patients have to undertake conversion between these two drugs. Additionally, it benefits the workflow in the clinical laboratory. The study was published in the August 2017 issue of the journal Practical Laboratory Medicine.
Latest Immunology News
- Stem Cell Test Predicts Treatment Outcome for Patients with Platinum-Resistant Ovarian Cancer
- Machine Learning-Enabled Blood Test Predicts Immunotherapy Response in Lymphoma Patients
- Post-Treatment Blood Test Could Inform Future Cancer Therapy Decisions
- Cerebrospinal Fluid Test Predicts Dangerous Side Effect of Cancer Treatment
- New Test Measures Preterm Infant Immunity Using Only Two Drops of Blood
- Simple Blood Test Could Help Choose Better Treatments for Patients with Recurrent Endometrial Cancer
- Novel Analytical Method Tracks Progression of Autoimmune Diseases
- 3D Bioprinted Gastric Cancer Model Uses Patient-Derived Tissue Fragments to Predict Drug Response
- Blood Test for Fungal Infections Could End Invasive Tissue Biopsies
- Cutting-Edge Microscopy Technology Enables Tailored Rheumatology Therapies
- New Discovery in Blood Immune Cells Paves Way for Parkinson's Disease Diagnostic Test
- AI Tool Uses Routine Blood Tests to Predict Immunotherapy Response for Various Cancers
- Blood Test Can Predict How Long Vaccine Immunity Will Last
- Microfluidic Chip-Based Device to Measure Viral Immunity
- Simple Blood Test Could Detect Drug Resistance in Ovarian Cancer Patients
Channels
Clinical Chemistry
view channel
Carbon Nanotubes Help Build Highly Accurate Sensors for Continuous Health Monitoring
Current sensors can measure various health indicators, such as blood glucose levels, in the body. However, there is a need to develop more accurate and sensitive sensor materials that can detect lower... Read more
Paper-Based Device Boosts HIV Test Accuracy from Dried Blood Samples
In regions where access to clinics for routine blood tests presents financial and logistical obstacles, HIV patients are increasingly able to collect and send a drop of blood using paper-based devices... Read moreMolecular Diagnostics
view channel
RNA-Based Blood Test Detects Preeclampsia Risk Months Before Symptoms
Preeclampsia remains a major cause of maternal morbidity and mortality, as well as preterm births. Despite current guidelines that aim to identify pregnant women at increased risk of preeclampsia using... Read more
First Of Its Kind Test Uses microRNAs to Predict Toxicity from Cancer Therapy
Many men with early-stage prostate cancer receive stereotactic body radiotherapy (SBRT), a highly precise form of radiation treatment that is completed in just five sessions. Compared to traditional radiation,... Read more
Novel Cell-Based Assay Provides Sensitive and Specific Autoantibody Detection in Demyelination
Anti-myelin-associated glycoprotein (MAG) antibodies serve as markers for an autoimmune demyelinating disorder that affects the peripheral nervous system, leading to sensory impairment. Anti-MAG-IgM antibodies... Read moreHematology
view channel
New Scoring System Predicts Risk of Developing Cancer from Common Blood Disorder
Clonal cytopenia of undetermined significance (CCUS) is a blood disorder commonly found in older adults, characterized by mutations in blood cells and a low blood count, but without any obvious cause or... Read more
Non-Invasive Prenatal Test for Fetal RhD Status Demonstrates 100% Accuracy
In the United States, approximately 15% of pregnant individuals are RhD-negative. However, in about 40% of these cases, the fetus is also RhD-negative, making the administration of RhoGAM unnecessary.... Read moreMicrobiology
view channel
Handheld Device Deliver Low-Cost TB Results in Less Than One Hour
Tuberculosis (TB) remains the deadliest infectious disease globally, affecting an estimated 10 million people annually. In 2021, about 4.2 million TB cases went undiagnosed or unreported, mainly due to... Read more
New AI-Based Method Improves Diagnosis of Drug-Resistant Infections
Drug-resistant infections, particularly those caused by deadly bacteria like tuberculosis and staphylococcus, are rapidly emerging as a global health emergency. These infections are more difficult to treat,... Read more
Breakthrough Diagnostic Technology Identifies Bacterial Infections with Almost 100% Accuracy within Three Hours
Rapid and precise identification of pathogenic microbes in patient samples is essential for the effective treatment of acute infectious diseases, such as sepsis. The fluorescence in situ hybridization... Read morePathology
view channel
Advanced Imaging Reveals Mechanisms Causing Autoimmune Disease
Myasthenia gravis, an autoimmune disease, leads to muscle weakness that can affect a range of muscles, including those needed for basic actions like blinking, smiling, or moving. Researchers have long... Read more
AI Model Effectively Predicts Patient Outcomes in Common Lung Cancer Type
Lung adenocarcinoma, the most common form of non-small cell lung cancer (NSCLC), typically adopts one of six distinct growth patterns, often combining multiple patterns within a single tumor.... Read moreTechnology
view channel
Pain-On-A-Chip Microfluidic Device Determines Types of Chronic Pain from Blood Samples
Chronic pain is a widespread condition that remains difficult to manage, and existing clinical methods for its treatment rely largely on self-reporting, which can be subjective and especially problematic... Read more
Innovative, Label-Free Ratiometric Fluorosensor Enables More Sensitive Viral RNA Detection
Viruses present a major global health risk, as demonstrated by recent pandemics, making early detection and identification essential for preventing new outbreaks. While traditional detection methods are... Read moreIndustry
view channel
Cepheid and Oxford Nanopore Technologies Partner on Advancing Automated Sequencing-Based Solutions
Cepheid (Sunnyvale, CA, USA), a leading molecular diagnostics company, and Oxford Nanopore Technologies (Oxford, UK), the company behind a new generation of sequencing-based molecular analysis technologies,... Read more
Grifols and Tecan’s IBL Collaborate on Advanced Biomarker Panels
Grifols (Barcelona, Spain), one of the world’s leading producers of plasma-derived medicines and innovative diagnostic solutions, is expanding its offer in clinical diagnostics through a strategic partnership... Read more