Analytical Performance Evaluated for Troponin T Gen 5 Assay

By LabMedica International staff writers
Posted on 05 Sep 2019
Print article
Image: The Elecsys Troponin T Gen 5 STAT assay (Photo courtesy of Roche Diagnostics).
Image: The Elecsys Troponin T Gen 5 STAT assay (Photo courtesy of Roche Diagnostics).
Increased blood concentrations of cardiac troponin (cTn) I and T are indicative of myocardial injury and are the preferred biomarkers for diagnosing acute myocardial infarction (AMI). The diagnosis of AMI requires a rise and/or fall in serial cTn measurements, in addition to other clinical features.

The introduction of high-sensitivity assays has enabled the rapid detection of subtle changes in cTn and facilitated earlier decision-making in the management of patients with suspected AMI, compared with less sensitive contemporary assays. These newer assays which can measure cTn concentrations 5 to 10 times lower than current assays and do so with improved analytical imprecision.

A team of Medical Scientists collaborating with the University of California San Diego Health (San Diego, CA, USA) analyzed the performance of the Elecsys Troponin T Gen 5 STAT assay. This assay is an electrochemiluminescence sandwich immunoassay, which uses two antibodies to form a sandwich complex with cTnT. A biotinylated monoclonal anti-cTnT-specific antibody and a monoclonal anti-cTnT-specific antibody labeled with ruthenium react to form a sandwich complex. Precision was evaluated per Clinical and Laboratory Standards Institute (CLSI) EP05-A2 using lithium-heparin plasma/quality control samples on Roche cobas e 411/cobas e 601 analyzers.

The scientists reported that the coefficients of variation (CV) for repeatability/intermediate precision were 0.7–5.6%/1.4–10.3% (cobas e 411; mean cardiac troponin T [cTnT]: 7.3–9,341 ng/L) and 0.7–3.0%/1.5–6.4% (cobas e 601; mean cTnT: 7.4–9,455 ng/L). There was no cross-reactivity with skeletal muscle troponin T (≤ 10,000 ng/L), skeletal muscle troponin I (≤ 100,000 ng/L), cardiac troponin I (≤ 10,000 ng/L), or human troponin C (≤ 80,000 ng/L). No interference was observed with biotin (≤ 20 ng/mL) or 34 drugs. The specified limit of blank and limit of detection were 3 ng/L and 5 ng/L, respectively, on the cobas e 411 analyzer, and 2.5 ng/L and 3 ng/L, respectively, on the cobas e 601 analyzer.

The authors concluded that the Elecsys TnT Gen 5 STAT assay demonstrated good analytical performance on cobas e 411 and cobas e 601 analyzers, with a CV of ≤ 10% at the 99th percentile URL. These findings support the routine use of the assay in clinical laboratories in the USA. The study was published in the August 2019 issue of the journal Clinica Chimica Acta.

Related Links:
University of California San Diego Health

Platinum Member
COVID-19 Rapid Test
OSOM COVID-19 Antigen Rapid Test
Magnetic Bead Separation Modules
MAG and HEATMAG
Anti-Cyclic Citrullinated Peptide Test
GPP-100 Anti-CCP Kit
New
Gold Member
Liquid Ready-To-Use Lp(a) Reagent
Lipoprotein (a) Reagent

Print article

Channels

Hematology

view channel
Image: The CAPILLARYS 3 DBS devices have received U.S. FDA 510(k) clearance (Photo courtesy of Sebia)

Next Generation Instrument Screens for Hemoglobin Disorders in Newborns

Hemoglobinopathies, the most widespread inherited conditions globally, affect about 7% of the population as carriers, with 2.7% of newborns being born with these conditions. The spectrum of clinical manifestations... Read more

Immunology

view channel
Image: The AI predictive model identifies the most potent cancer killing immune cells for use in immunotherapies (Photo courtesy of Shutterstock)

AI Predicts Tumor-Killing Cells with High Accuracy

Cellular immunotherapy involves extracting immune cells from a patient's tumor, potentially enhancing their cancer-fighting capabilities through engineering, and then expanding and reintroducing them into the body.... Read more

Microbiology

view channel
Image: The T-SPOT.TB test is now paired with the Auto-Pure 2400 liquid handling platform for accurate TB testing (Photo courtesy of Shutterstock)

Integrated Solution Ushers New Era of Automated Tuberculosis Testing

Tuberculosis (TB) is responsible for 1.3 million deaths every year, positioning it as one of the top killers globally due to a single infectious agent. In 2022, around 10.6 million people were diagnosed... Read more