The analytical performance of an immunoenzymometric assay for the cardiac troponin (cTnI) has been appraised and compared with other methods.

The test is a two-site immunoenzymometric assay, which uses a combination of two monoclonal antibodies, respectively directed to 41–49 and 87–91 amino acids of the cTnI peptide chain, and the ternary troponin ITC complex as a calibration antigen.

Scientists at the Fondazione Toscana G. Monasterio (Pisa, Italy) evaluated the assay in 452 healthy individuals of which 326 were males and 126 were female. The median age was 45 years with a range between 17 to 76 years. The study participants were recruited from laboratory staff, blood donors, or voluntary subjects, included in screening programs for preventive medicine.

The immunoenzymometric assay for the cTnI evaluated in the study was the Tosoh ST AIA-PACK cTnI 3rd-Generation, and it uses the automated AIA-2000 platform (Tosoh Corporation, Tokyo, Japan). The third-generation AIA-Pack assay for cTnI showed an improved analytical sensitivity and reproducibility, especially at very low cTnI concentrations compared to the previous second-generation AIA-Pack assay. The values of the third-generation assay were a limit of detection (LoD) at 8.7 ng/L and the limit of quantitation (LoQ) at 100 ng/L. The LoD and 10% LoQ values of the second-generation assay reported previously were 38 ng/L and 130 ng/L, respectively.

The third-generation AIA-Pack Tosoh assay for cTnI showed a very close agreement throughout all the working range with the Access AccuTnI Beckman–Coulter method (Beckman Coulter, Inc.; Fullerton, CA, USA). Close concordance was also demonstrated between the cTnI measured by Tosoh methods and those of cTnT measured with ECLIA Roche method (Roche Diagnostics, Mannheim, Germany). The authors concluded that their results imply that the third generation AIA-Pack assay is suitable for the clinical evaluation of patients with cardiac diseases. The study was published in the January 2013 issue of the journal Clinica Chimica Acta.

Related Links:
Fondazione Toscana G. Monasterio
Beckman Coulter Inc.
Roche Diagnostics

Platinum Member
COVID-19 Rapid Test
OSOM COVID-19 Antigen Rapid Test
Magnetic Bead Separation Modules
MAG and HEATMAG
POCT Fluorescent Immunoassay Analyzer
FIA Go
New
Gold Member
TORCH Panel Rapid Test
Rapid TORCH Panel Test

Print article

Channels

Hematology

view channel
Image: The CAPILLARYS 3 DBS devices have received U.S. FDA 510(k) clearance (Photo courtesy of Sebia)

Next Generation Instrument Screens for Hemoglobin Disorders in Newborns

Hemoglobinopathies, the most widespread inherited conditions globally, affect about 7% of the population as carriers, with 2.7% of newborns being born with these conditions. The spectrum of clinical manifestations... Read more

Immunology

view channel
Image: The AI predictive model identifies the most potent cancer killing immune cells for use in immunotherapies (Photo courtesy of Shutterstock)

AI Predicts Tumor-Killing Cells with High Accuracy

Cellular immunotherapy involves extracting immune cells from a patient's tumor, potentially enhancing their cancer-fighting capabilities through engineering, and then expanding and reintroducing them into the body.... Read more

Microbiology

view channel
Image: The T-SPOT.TB test is now paired with the Auto-Pure 2400 liquid handling platform for accurate TB testing (Photo courtesy of Shutterstock)

Integrated Solution Ushers New Era of Automated Tuberculosis Testing

Tuberculosis (TB) is responsible for 1.3 million deaths every year, positioning it as one of the top killers globally due to a single infectious agent. In 2022, around 10.6 million people were diagnosed... Read more