Analytical Performance Evaluated for Troponin T Gen 5 Assay

By LabMedica International staff writers
Posted on 05 Sep 2019
Print article
Image: The Elecsys Troponin T Gen 5 STAT assay (Photo courtesy of Roche Diagnostics).
Image: The Elecsys Troponin T Gen 5 STAT assay (Photo courtesy of Roche Diagnostics).
Increased blood concentrations of cardiac troponin (cTn) I and T are indicative of myocardial injury and are the preferred biomarkers for diagnosing acute myocardial infarction (AMI). The diagnosis of AMI requires a rise and/or fall in serial cTn measurements, in addition to other clinical features.

The introduction of high-sensitivity assays has enabled the rapid detection of subtle changes in cTn and facilitated earlier decision-making in the management of patients with suspected AMI, compared with less sensitive contemporary assays. These newer assays which can measure cTn concentrations 5 to 10 times lower than current assays and do so with improved analytical imprecision.

A team of Medical Scientists collaborating with the University of California San Diego Health (San Diego, CA, USA) analyzed the performance of the Elecsys Troponin T Gen 5 STAT assay. This assay is an electrochemiluminescence sandwich immunoassay, which uses two antibodies to form a sandwich complex with cTnT. A biotinylated monoclonal anti-cTnT-specific antibody and a monoclonal anti-cTnT-specific antibody labeled with ruthenium react to form a sandwich complex. Precision was evaluated per Clinical and Laboratory Standards Institute (CLSI) EP05-A2 using lithium-heparin plasma/quality control samples on Roche cobas e 411/cobas e 601 analyzers.

The scientists reported that the coefficients of variation (CV) for repeatability/intermediate precision were 0.7–5.6%/1.4–10.3% (cobas e 411; mean cardiac troponin T [cTnT]: 7.3–9,341 ng/L) and 0.7–3.0%/1.5–6.4% (cobas e 601; mean cTnT: 7.4–9,455 ng/L). There was no cross-reactivity with skeletal muscle troponin T (≤ 10,000 ng/L), skeletal muscle troponin I (≤ 100,000 ng/L), cardiac troponin I (≤ 10,000 ng/L), or human troponin C (≤ 80,000 ng/L). No interference was observed with biotin (≤ 20 ng/mL) or 34 drugs. The specified limit of blank and limit of detection were 3 ng/L and 5 ng/L, respectively, on the cobas e 411 analyzer, and 2.5 ng/L and 3 ng/L, respectively, on the cobas e 601 analyzer.

The authors concluded that the Elecsys TnT Gen 5 STAT assay demonstrated good analytical performance on cobas e 411 and cobas e 601 analyzers, with a CV of ≤ 10% at the 99th percentile URL. These findings support the routine use of the assay in clinical laboratories in the USA. The study was published in the August 2019 issue of the journal Clinica Chimica Acta.

Related Links:
University of California San Diego Health

Gold Member
Troponin T QC
Troponin T Quality Control
Verification Panels for Assay Development & QC
Seroconversion Panels
New
Silver Member
ACTH Assay
ACTH ELISA
New
Respiratory Bacterial Panel
Real Respiratory Bacterial Panel 2

Print article

Channels

Molecular Diagnostics

view channel
Image: The study investigated D-dimer testing in patients who are at higher risk of pulmonary embolism (Photo courtesy of Adobe Stock)

D-Dimer Testing Can Identify Patients at Higher Risk of Pulmonary Embolism

Pulmonary embolism (PE) is a commonly suspected condition in emergency departments (EDs) and can be life-threatening if not diagnosed correctly. Achieving an accurate diagnosis is vital for providing effective... Read more

Immunology

view channel
Image: The findings were based on patients from the ADAURA clinical trial of the targeted therapy osimertinib for patients with NSCLC with EGFR-activated mutations (Photo courtesy of YSM Multimedia Team)

Post-Treatment Blood Test Could Inform Future Cancer Therapy Decisions

In the ongoing advancement of personalized medicine, a new study has provided evidence supporting the use of a tool that detects cancer-derived molecules in the blood of lung cancer patients years after... Read more

Microbiology

view channel
Image: Schematic representation illustrating the key findings of the study (Photo courtesy of UNIST)

Breakthrough Diagnostic Technology Identifies Bacterial Infections with Almost 100% Accuracy within Three Hours

Rapid and precise identification of pathogenic microbes in patient samples is essential for the effective treatment of acute infectious diseases, such as sepsis. The fluorescence in situ hybridization... Read more
Sekisui Diagnostics UK Ltd.