C-Reactive Protein Point-Of-Care Test Evaluated
By LabMedica International staff writers Posted on 27 Aug 2019 |
Image: The cobas b 101 POC system; an in vitro diagnostic (IVD) test system offering C-reactive protein, HbA1c and a complete lipid profile (CHOL, HDL, LDL, TG) on one device at the point of care (Photo courtesy of Roche Diagnostics).
C-reactive protein (CRP) is a key mediator of the acute-phase response, with blood levels of CRP increasing rapidly after an inflammatory stimulus. Therefore, changes in serum levels of CRP are a clinically useful marker of infection, inflammation, and tissue injury.
CRP testing in the primary setting as in point-of-care (POC) can help reduce diagnostic uncertainty by differentiating between bacterial and viral infections and has been shown to be cost-effective for reducing inappropriate antibiotic prescriptions. However, despite the availability of POC devices, POC CRP testing has not yet been widely adopted in primary care clinics.
Clinical laboratory scientists at the Catharina Hospital and Technical University Eindhoven (Eindhoven, The Netherland) and their colleagues evaluated a POC CPR test at three sites and one reference laboratory. Within-run (repeatability), within-laboratory (intermediate precision), and between-laboratory precision (reproducibility) were assessed. Method comparison and matrix/lot-to-lot comparison studies were conducted using prospectively collected blood samples from 217 adults (apparently healthy or with clinically relevant conditions).
The investigators compared the analytical performance of the cobas b 101 POC system, which provides glycated hemoglobin (HbA1c) and lipid panel tests (measurement of cholesterol, triglyceride, and high-density lipoprotein; calculation of low-density lipoprotein) for managing diabetes and dyslipidemia at point-of-need. This method was compared with a reference test: CRPNX reagent on a Roche cobas c 501 module.
The scientists reported that clinically relevant CRP concentrations measured with the CRP Test showed good agreement with those measured by CRPNX reagent. Coefficients of variation (CV) for repeatability and intermediate precision ranged from 1.7%–4.0% and 1.9%–4.5%, respectively, for human serum pools containing CRP 4.7–350.7 mg/L; repeatability in clinical samples ranged from 1.6%–5.9% (3.3–360.3 mg/L). CVs for reproducibility ranged from 2.5%–4.0% (4.7–344.3 mg/L). CRP concentrations were comparable for capillary whole blood, serum, Li-heparin whole blood/plasma, K2 and K3 EDTA whole blood/plasma. The overall mean usability score was 4.18/5 and the error rate across 9,378 tests was 1.00%.
The authors concluded that their findings indicate that healthcare professionals can obtain precise and reproducible CRP values with the cobas POC CRP Test that show very good correlation with laboratory measurements. Importantly, operators considered the system convenient for use in the POC environment. The study was published in the September 2019 issue of the journal Clinical Biochemistry.
Related Links:
Catharina Hospital and Technical University Eindhoven
CRP testing in the primary setting as in point-of-care (POC) can help reduce diagnostic uncertainty by differentiating between bacterial and viral infections and has been shown to be cost-effective for reducing inappropriate antibiotic prescriptions. However, despite the availability of POC devices, POC CRP testing has not yet been widely adopted in primary care clinics.
Clinical laboratory scientists at the Catharina Hospital and Technical University Eindhoven (Eindhoven, The Netherland) and their colleagues evaluated a POC CPR test at three sites and one reference laboratory. Within-run (repeatability), within-laboratory (intermediate precision), and between-laboratory precision (reproducibility) were assessed. Method comparison and matrix/lot-to-lot comparison studies were conducted using prospectively collected blood samples from 217 adults (apparently healthy or with clinically relevant conditions).
The investigators compared the analytical performance of the cobas b 101 POC system, which provides glycated hemoglobin (HbA1c) and lipid panel tests (measurement of cholesterol, triglyceride, and high-density lipoprotein; calculation of low-density lipoprotein) for managing diabetes and dyslipidemia at point-of-need. This method was compared with a reference test: CRPNX reagent on a Roche cobas c 501 module.
The scientists reported that clinically relevant CRP concentrations measured with the CRP Test showed good agreement with those measured by CRPNX reagent. Coefficients of variation (CV) for repeatability and intermediate precision ranged from 1.7%–4.0% and 1.9%–4.5%, respectively, for human serum pools containing CRP 4.7–350.7 mg/L; repeatability in clinical samples ranged from 1.6%–5.9% (3.3–360.3 mg/L). CVs for reproducibility ranged from 2.5%–4.0% (4.7–344.3 mg/L). CRP concentrations were comparable for capillary whole blood, serum, Li-heparin whole blood/plasma, K2 and K3 EDTA whole blood/plasma. The overall mean usability score was 4.18/5 and the error rate across 9,378 tests was 1.00%.
The authors concluded that their findings indicate that healthcare professionals can obtain precise and reproducible CRP values with the cobas POC CRP Test that show very good correlation with laboratory measurements. Importantly, operators considered the system convenient for use in the POC environment. The study was published in the September 2019 issue of the journal Clinical Biochemistry.
Related Links:
Catharina Hospital and Technical University Eindhoven
Latest Clinical Chem. News
- 3D Printed Point-Of-Care Mass Spectrometer Outperforms State-Of-The-Art Models
- POC Biomedical Test Spins Water Droplet Using Sound Waves for Cancer Detection
- Highly Reliable Cell-Based Assay Enables Accurate Diagnosis of Endocrine Diseases
- New Blood Testing Method Detects Potent Opioids in Under Three Minutes
- Wireless Hepatitis B Test Kit Completes Screening and Data Collection in One Step
- Pain-Free, Low-Cost, Sensitive, Radiation-Free Device Detects Breast Cancer in Urine
- Spit Test Detects Breast Cancer in Five Seconds
- Electrochemical Sensors with Next-Generation Coating Advances Precision Diagnostics at POC
- First-Of-Its-Kind Handheld Device Accurately Detects Fentanyl in Urine within Seconds
- New Fluorescent Sensor Array Lights up Alzheimer’s-Related Proteins for Earlier Detection
- Automated Mass Spectrometry-Based Clinical Analyzer Could Transform Lab Testing
- Highly Sensitive pH Sensor to Aid Detection of Cancers and Vector-Borne Viruses
- Non-Invasive Sensor Monitors Changes in Saliva Compositions to Rapidly Diagnose Diabetes
- Breakthrough Immunoassays to Aid in Risk Assessment of Preeclampsia
- Urine Test for Monitoring Changes in Kidney Health Markers Can Predict New-Onset Heart Failure
- AACC Releases Comprehensive Diabetes Testing Guidelines