Novel Turbidimetric Immunoassay Evaluated for Fecal Calprotectin
By LabMedica International staff writers Posted on 28 Sep 2016 |
Image: The BS-380 automatic biochemistry analyzer (Photo courtesy of Mindray).
Calprotectin is a multifunctional protein that plays an important role in the diagnosis and follow-up of inflammatory bowel disease and high levels of calprotectin in stool samples are associated with inflammation of the intestinal tract.
Fecal calprotectin assays are widely used to exclude inflammatory bowel disease (IBD) in patients with suspected IBD, but the problem with most of the fecal calprotectin assays is the rather long test-turnaround times, before the results are available to the physician.
Scientists at Uppsala University (Sweden) optimized a particle enhanced turbidimetric immunoassay (PETIA) for fecal calprotectin and validated the assay for two clinical autoanalyzers using routine fecal samples. They compared the PETIA with a commercial enzyme-linked immunosorbent assay (ELISA), known as the Bühlmann fCAL ELISA (Bühlmann Laboratories, Schönenbuch, Switzerland). This new latex based turbidimetric calprotectin assay applies particles coated with anti-human calprotectin (MRP8/14) antibodies: the agglutination is proportional to the calprotectin concentration.
The fecal calprotectin PETIA was validated on two chemistry analyzers, the Mindray BS-380 (Mindray, Shenzhen, China) and the Cobas 501 (Roche Diagnostics, Basel, Switzerland). The assay is linear in the range 11 μg/g to 2,000 μg/g, with a limit of quantitation of approximately 10 μg/g. No antigen excess hook effect was observed up to 10,000 μg/g to 15,000 μg/g depending on the instrument used. The turbidimetric method showed a good agreement with the Bühlmann ELISA. The total coefficient of variation was 3% to 8% in the 50 μg/g to100 μg/g range.
The authors concluded that the fecal calprotectin PETIA, fCal Turbo, is well suited for rapid analysis of fecal calprotectin on Mindray BS-380 or Cobas 501 clinical chemistry analyzers. The test results are concordant with the Bühlmann fecal MRP8/14 ELISA. The study was published in the September 2016 issue of the Journal of Clinical Laboratory Analysis.
Related Links:
Uppsala University
Bühlmann Laboratories
Mindray
Roche Diagnostics
Fecal calprotectin assays are widely used to exclude inflammatory bowel disease (IBD) in patients with suspected IBD, but the problem with most of the fecal calprotectin assays is the rather long test-turnaround times, before the results are available to the physician.
Scientists at Uppsala University (Sweden) optimized a particle enhanced turbidimetric immunoassay (PETIA) for fecal calprotectin and validated the assay for two clinical autoanalyzers using routine fecal samples. They compared the PETIA with a commercial enzyme-linked immunosorbent assay (ELISA), known as the Bühlmann fCAL ELISA (Bühlmann Laboratories, Schönenbuch, Switzerland). This new latex based turbidimetric calprotectin assay applies particles coated with anti-human calprotectin (MRP8/14) antibodies: the agglutination is proportional to the calprotectin concentration.
The fecal calprotectin PETIA was validated on two chemistry analyzers, the Mindray BS-380 (Mindray, Shenzhen, China) and the Cobas 501 (Roche Diagnostics, Basel, Switzerland). The assay is linear in the range 11 μg/g to 2,000 μg/g, with a limit of quantitation of approximately 10 μg/g. No antigen excess hook effect was observed up to 10,000 μg/g to 15,000 μg/g depending on the instrument used. The turbidimetric method showed a good agreement with the Bühlmann ELISA. The total coefficient of variation was 3% to 8% in the 50 μg/g to100 μg/g range.
The authors concluded that the fecal calprotectin PETIA, fCal Turbo, is well suited for rapid analysis of fecal calprotectin on Mindray BS-380 or Cobas 501 clinical chemistry analyzers. The test results are concordant with the Bühlmann fecal MRP8/14 ELISA. The study was published in the September 2016 issue of the Journal of Clinical Laboratory Analysis.
Related Links:
Uppsala University
Bühlmann Laboratories
Mindray
Roche Diagnostics
Latest Clinical Chem. News
- 3D Printed Point-Of-Care Mass Spectrometer Outperforms State-Of-The-Art Models
- POC Biomedical Test Spins Water Droplet Using Sound Waves for Cancer Detection
- Highly Reliable Cell-Based Assay Enables Accurate Diagnosis of Endocrine Diseases
- New Blood Testing Method Detects Potent Opioids in Under Three Minutes
- Wireless Hepatitis B Test Kit Completes Screening and Data Collection in One Step
- Pain-Free, Low-Cost, Sensitive, Radiation-Free Device Detects Breast Cancer in Urine
- Spit Test Detects Breast Cancer in Five Seconds
- Electrochemical Sensors with Next-Generation Coating Advances Precision Diagnostics at POC
- First-Of-Its-Kind Handheld Device Accurately Detects Fentanyl in Urine within Seconds
- New Fluorescent Sensor Array Lights up Alzheimer’s-Related Proteins for Earlier Detection
- Automated Mass Spectrometry-Based Clinical Analyzer Could Transform Lab Testing
- Highly Sensitive pH Sensor to Aid Detection of Cancers and Vector-Borne Viruses
- Non-Invasive Sensor Monitors Changes in Saliva Compositions to Rapidly Diagnose Diabetes
- Breakthrough Immunoassays to Aid in Risk Assessment of Preeclampsia
- Urine Test for Monitoring Changes in Kidney Health Markers Can Predict New-Onset Heart Failure
- AACC Releases Comprehensive Diabetes Testing Guidelines