Hyaluronic Acid Turbidimetric Assay Compared to Standard Method
By LabMedica International staff writers Posted on 13 Jan 2016 |
Image: The Hitachi 917 Automatic Disk-Chemistry Analyzer (Photo courtesy of Roche Diagnostics).
Circulating hyaluronic acid (HA) in human adults is primarily produced in the peripheral soft connective tissue and transported to the systemic circulation via lymph drainage and the majority of HA is removed from circulation by hepatic elimination.
HA is essentially non-immunogenic, which has excluded direct immunochemical methods of measurement. To accommodate this, several advanced methods of measurements have been used including enzymatic degradation; hyaluronic-binding protein (HABP) linked enzyme-linked immunosorbent assay (ELISA) and high-performance liquid chromatography.
Scientists at the Nordsjaellands Hospital, University of Copenhagen, (Denmark) and their colleagues measured HA concentrations in 39 samples of serum from 39 randomly selected intensive care unit (ICU) patients. The HA was measured by a particle-enhanced turbidimetric immunoassay (PETIA) and enzyme-linked immunosorbent assay (ELISA) in a 40-sample dilution series and the 39 ICU patients.
The HA was measured with the PETIA (Corgenix; Broomfield, CO, USA) in ICU samples on the Hitachi 917 (Roche Diagnostics, GmbH; Mannheim, Germany) and in the dilution series on Modular P (Roche Diagnostics GmbH) analyzers (test assay), and by Corgenix HA ELISA using double determination (reference method).
The scientists found that in the ICU patients, the median HA concentration was 159.0 ng/mL (interquartile range (IQR) 117.5–362.5 ng/mL) with ELISA and 157.5 ng/ml (IQR 92.5–359.6 ng/mL) with PETIA. The mean difference was 12.88 ng/mL which was statistically significant and the 95% limits of agreement were −91.17 to 116.9 ng/mL. In the dilution series, the mean difference was −59.26 ng/mL (95% CI, −74.68 to 43.84 ng/mL, and the 95% limits of agreement were 35.23 to −153.8 ng/mL.
The authors concluded that there was random variation between the PETIA and ELISA test that could affect performance in a clinical context. The new clinical biochemistry assay for HA determination will allow for large studies of the clinical utility of HA. The study was published online on December 14, 2015, in the Journal of Clinical Laboratory Analysis.
Related Links:
Nordsjaellands Hospital
Corgenix
Roche Diagnostics GmbH
HA is essentially non-immunogenic, which has excluded direct immunochemical methods of measurement. To accommodate this, several advanced methods of measurements have been used including enzymatic degradation; hyaluronic-binding protein (HABP) linked enzyme-linked immunosorbent assay (ELISA) and high-performance liquid chromatography.
Scientists at the Nordsjaellands Hospital, University of Copenhagen, (Denmark) and their colleagues measured HA concentrations in 39 samples of serum from 39 randomly selected intensive care unit (ICU) patients. The HA was measured by a particle-enhanced turbidimetric immunoassay (PETIA) and enzyme-linked immunosorbent assay (ELISA) in a 40-sample dilution series and the 39 ICU patients.
The HA was measured with the PETIA (Corgenix; Broomfield, CO, USA) in ICU samples on the Hitachi 917 (Roche Diagnostics, GmbH; Mannheim, Germany) and in the dilution series on Modular P (Roche Diagnostics GmbH) analyzers (test assay), and by Corgenix HA ELISA using double determination (reference method).
The scientists found that in the ICU patients, the median HA concentration was 159.0 ng/mL (interquartile range (IQR) 117.5–362.5 ng/mL) with ELISA and 157.5 ng/ml (IQR 92.5–359.6 ng/mL) with PETIA. The mean difference was 12.88 ng/mL which was statistically significant and the 95% limits of agreement were −91.17 to 116.9 ng/mL. In the dilution series, the mean difference was −59.26 ng/mL (95% CI, −74.68 to 43.84 ng/mL, and the 95% limits of agreement were 35.23 to −153.8 ng/mL.
The authors concluded that there was random variation between the PETIA and ELISA test that could affect performance in a clinical context. The new clinical biochemistry assay for HA determination will allow for large studies of the clinical utility of HA. The study was published online on December 14, 2015, in the Journal of Clinical Laboratory Analysis.
Related Links:
Nordsjaellands Hospital
Corgenix
Roche Diagnostics GmbH
Latest Clinical Chem. News
- 3D Printed Point-Of-Care Mass Spectrometer Outperforms State-Of-The-Art Models
- POC Biomedical Test Spins Water Droplet Using Sound Waves for Cancer Detection
- Highly Reliable Cell-Based Assay Enables Accurate Diagnosis of Endocrine Diseases
- New Blood Testing Method Detects Potent Opioids in Under Three Minutes
- Wireless Hepatitis B Test Kit Completes Screening and Data Collection in One Step
- Pain-Free, Low-Cost, Sensitive, Radiation-Free Device Detects Breast Cancer in Urine
- Spit Test Detects Breast Cancer in Five Seconds
- Electrochemical Sensors with Next-Generation Coating Advances Precision Diagnostics at POC
- First-Of-Its-Kind Handheld Device Accurately Detects Fentanyl in Urine within Seconds
- New Fluorescent Sensor Array Lights up Alzheimer’s-Related Proteins for Earlier Detection
- Automated Mass Spectrometry-Based Clinical Analyzer Could Transform Lab Testing
- Highly Sensitive pH Sensor to Aid Detection of Cancers and Vector-Borne Viruses
- Non-Invasive Sensor Monitors Changes in Saliva Compositions to Rapidly Diagnose Diabetes
- Breakthrough Immunoassays to Aid in Risk Assessment of Preeclampsia
- Urine Test for Monitoring Changes in Kidney Health Markers Can Predict New-Onset Heart Failure
- AACC Releases Comprehensive Diabetes Testing Guidelines