Low Density Lipoprotein Apolipoprotein-B Assay Validated
By LabMedica International staff writers Posted on 19 Nov 2014 |
The relationship between apolipoprotein-B (Apo-B) concentration and the risk of developing cardiovascular disease (CVD) is more robust than the relationship between low density lipoprotein cholesterol (LDL-C) concentration and CVD.
It is however unknown if Apo-B or low density lipoprotein (LDL) particle number (LDL-P) has a stronger predictive ability for future CHD and this uncertainty may be due to the inability of current Apo-B assays to separate the contribution of very low-density lipoprotein particles (VLDL-P) from the total Apo-B concentration.
Scientists at the Dartmouth-Hitchcock Medical Center (Lebanon, NH, USA) obtained a total of 300 lithium heparinized patient plasma samples submitted for routine fasting lipid profile. The profile included total cholesterol, high density lipoprotein (HDL), triglycerides and calculated LDL cholesterol and was included for analysis via both the LDL Apo-B assay as well as a Direct LDL assay on a specific immunoassay analyzer.
The LDL Apo-B assay (Maine Standards Company; Windham, ME, USA) for use on the Roche Cobas 6000 analyzer (Indianapolis, IN, USA) is an immunoturbidometric assay employing a polyclonal antiserum against Apo-B. In order to accurately quantify only the Apo-B content on LDL particles, the patient samples are first automatically diluted via the instrument. The collaborating scientists at Maine Standards developed an automated immunoturbidimetric assay for the Cobas 6000 analyzer that specifically quantifies the Apo-B present on the surface of LDL particles.
The LDL Apo-B within-run imprecision was 2.3% at 62 mg/dL and 2.2% at 109 mg/dL. The within-laboratory imprecision was 9.7% at 57 mg/dl and 6.1% at 104 mg/dL. There was no strong linear relationship between the calculated or measured LDL cholesterol concentration and the measured LDL Apo-B concentration. The assay demonstrated laboratory acceptable precision, low limit of quantitation (LoQ) and a wide measurement range. This assay that specifically measures Apo-B on LDL particles demonstrated a positive correlation with LDL-C that is not unlike the relationship between LDL-C and total Apo-B.
The authors concluded that they had verified the relationship between the LDL Apo-B concentration and the LDL-particle assay. This precise and easy to use automated turbidimetric assay for use on the Roche Cobas 6000 analyzer may serve as either a low cost alternative to the high-complexity LDL-particle assay by nuclear magnetic resonance (NMR), or as a screen to identify which patients should have a follow up LDL-particle assessment performed in a reference laboratory. The study was published in the November 2014 issue of the journal Clinical Biochemistry.
Related Links:
Dartmouth-Hitchcock Medical Center
Maine Standards Company
Roche
It is however unknown if Apo-B or low density lipoprotein (LDL) particle number (LDL-P) has a stronger predictive ability for future CHD and this uncertainty may be due to the inability of current Apo-B assays to separate the contribution of very low-density lipoprotein particles (VLDL-P) from the total Apo-B concentration.
Scientists at the Dartmouth-Hitchcock Medical Center (Lebanon, NH, USA) obtained a total of 300 lithium heparinized patient plasma samples submitted for routine fasting lipid profile. The profile included total cholesterol, high density lipoprotein (HDL), triglycerides and calculated LDL cholesterol and was included for analysis via both the LDL Apo-B assay as well as a Direct LDL assay on a specific immunoassay analyzer.
The LDL Apo-B assay (Maine Standards Company; Windham, ME, USA) for use on the Roche Cobas 6000 analyzer (Indianapolis, IN, USA) is an immunoturbidometric assay employing a polyclonal antiserum against Apo-B. In order to accurately quantify only the Apo-B content on LDL particles, the patient samples are first automatically diluted via the instrument. The collaborating scientists at Maine Standards developed an automated immunoturbidimetric assay for the Cobas 6000 analyzer that specifically quantifies the Apo-B present on the surface of LDL particles.
The LDL Apo-B within-run imprecision was 2.3% at 62 mg/dL and 2.2% at 109 mg/dL. The within-laboratory imprecision was 9.7% at 57 mg/dl and 6.1% at 104 mg/dL. There was no strong linear relationship between the calculated or measured LDL cholesterol concentration and the measured LDL Apo-B concentration. The assay demonstrated laboratory acceptable precision, low limit of quantitation (LoQ) and a wide measurement range. This assay that specifically measures Apo-B on LDL particles demonstrated a positive correlation with LDL-C that is not unlike the relationship between LDL-C and total Apo-B.
The authors concluded that they had verified the relationship between the LDL Apo-B concentration and the LDL-particle assay. This precise and easy to use automated turbidimetric assay for use on the Roche Cobas 6000 analyzer may serve as either a low cost alternative to the high-complexity LDL-particle assay by nuclear magnetic resonance (NMR), or as a screen to identify which patients should have a follow up LDL-particle assessment performed in a reference laboratory. The study was published in the November 2014 issue of the journal Clinical Biochemistry.
Related Links:
Dartmouth-Hitchcock Medical Center
Maine Standards Company
Roche
Read the full article by registering today, it's FREE!
Register now for FREE to LabMedica.com and get complete access to news and events that shape the world of Clinical Laboratory Medicine.
- Free digital version edition of LabMedica International sent by email on regular basis
- Free print version of LabMedica International magazine (available only outside USA and Canada).
- Free and unlimited access to back issues of LabMedica International in digital format
- Free LabMedica International Newsletter sent every week containing the latest news
- Free breaking news sent via email
- Free access to Events Calendar
- Free access to LinkXpress new product services
- REGISTRATION IS FREE AND EASY!
Sign in: Registered website members
Sign in: Registered magazine subscribers
Latest Clinical Chem. News
- 3D Printed Point-Of-Care Mass Spectrometer Outperforms State-Of-The-Art Models
- POC Biomedical Test Spins Water Droplet Using Sound Waves for Cancer Detection
- Highly Reliable Cell-Based Assay Enables Accurate Diagnosis of Endocrine Diseases
- New Blood Testing Method Detects Potent Opioids in Under Three Minutes
- Wireless Hepatitis B Test Kit Completes Screening and Data Collection in One Step
- Pain-Free, Low-Cost, Sensitive, Radiation-Free Device Detects Breast Cancer in Urine
- Spit Test Detects Breast Cancer in Five Seconds
- Electrochemical Sensors with Next-Generation Coating Advances Precision Diagnostics at POC
- First-Of-Its-Kind Handheld Device Accurately Detects Fentanyl in Urine within Seconds
- New Fluorescent Sensor Array Lights up Alzheimer’s-Related Proteins for Earlier Detection
- Automated Mass Spectrometry-Based Clinical Analyzer Could Transform Lab Testing
- Highly Sensitive pH Sensor to Aid Detection of Cancers and Vector-Borne Viruses
- Non-Invasive Sensor Monitors Changes in Saliva Compositions to Rapidly Diagnose Diabetes
- Breakthrough Immunoassays to Aid in Risk Assessment of Preeclampsia
- Urine Test for Monitoring Changes in Kidney Health Markers Can Predict New-Onset Heart Failure
- AACC Releases Comprehensive Diabetes Testing Guidelines