Novel Miniature Mass Spectrometer Moves Cocaine Measurement to the Field
By LabMedica International staff writers Posted on 07 Jul 2014 |
A novel device for field testing of cocaine and other drugs comprises a portable mass spectrometer and a system for rapid drying and extracting urine samples.
To avoid the problem of transferring samples for drug analysis to distant central laboratories, investigators at the University of Toronto (Canada) designed a custom digital microfluidic system to deliver droplets of solvent to dried urine samples and then transport extracted analytes to an array of nanoelectrospray emitters for analysis. Tandem mass spectrometry (MS/MS) detection was performed using a fully autonomous 25 kg instrument.
Using this method, the investigators demonstrated that cocaine, benzoylecgonine, and codeine could be quantified from four samples in less than 15 minutes from (dried) sample to analysis. The limit of quantization (LOQ) for cocaine was 40 ng/mL, which is compatible with the performance criteria for laboratory analyses established by the United Nations Office on Drugs and Crime. The LOQ of the new method was superior to the 300 ng/mL cutoff values attained by immunoassay-based portable analysis systems.
The miniature MS system is attractive for the quantization of drugs of abuse from urine and, more generally, may be useful for a wide range of applications that would benefit from portable, quantitative, on-site analysis.
Related Links:
University of Toronto
To avoid the problem of transferring samples for drug analysis to distant central laboratories, investigators at the University of Toronto (Canada) designed a custom digital microfluidic system to deliver droplets of solvent to dried urine samples and then transport extracted analytes to an array of nanoelectrospray emitters for analysis. Tandem mass spectrometry (MS/MS) detection was performed using a fully autonomous 25 kg instrument.
Using this method, the investigators demonstrated that cocaine, benzoylecgonine, and codeine could be quantified from four samples in less than 15 minutes from (dried) sample to analysis. The limit of quantization (LOQ) for cocaine was 40 ng/mL, which is compatible with the performance criteria for laboratory analyses established by the United Nations Office on Drugs and Crime. The LOQ of the new method was superior to the 300 ng/mL cutoff values attained by immunoassay-based portable analysis systems.
The miniature MS system is attractive for the quantization of drugs of abuse from urine and, more generally, may be useful for a wide range of applications that would benefit from portable, quantitative, on-site analysis.
Related Links:
University of Toronto
Latest Clinical Chem. News
- 3D Printed Point-Of-Care Mass Spectrometer Outperforms State-Of-The-Art Models
- POC Biomedical Test Spins Water Droplet Using Sound Waves for Cancer Detection
- Highly Reliable Cell-Based Assay Enables Accurate Diagnosis of Endocrine Diseases
- New Blood Testing Method Detects Potent Opioids in Under Three Minutes
- Wireless Hepatitis B Test Kit Completes Screening and Data Collection in One Step
- Pain-Free, Low-Cost, Sensitive, Radiation-Free Device Detects Breast Cancer in Urine
- Spit Test Detects Breast Cancer in Five Seconds
- Electrochemical Sensors with Next-Generation Coating Advances Precision Diagnostics at POC
- First-Of-Its-Kind Handheld Device Accurately Detects Fentanyl in Urine within Seconds
- New Fluorescent Sensor Array Lights up Alzheimer’s-Related Proteins for Earlier Detection
- Automated Mass Spectrometry-Based Clinical Analyzer Could Transform Lab Testing
- Highly Sensitive pH Sensor to Aid Detection of Cancers and Vector-Borne Viruses
- Non-Invasive Sensor Monitors Changes in Saliva Compositions to Rapidly Diagnose Diabetes
- Breakthrough Immunoassays to Aid in Risk Assessment of Preeclampsia
- Urine Test for Monitoring Changes in Kidney Health Markers Can Predict New-Onset Heart Failure
- AACC Releases Comprehensive Diabetes Testing Guidelines