Prognostic Performance of High-Sensitivity Cardiac Troponin Assay Evaluated
By LabMedica International staff writers Posted on 28 Jan 2014 |
Image: the ARCHITECT clinical chemistry analyzer (Photo courtesy of Abbott).
Cardiac troponin is the preferred biomarker for diagnosis of acute myocardial infarction (MI) and is useful for risk stratification of patients with acute coronary syndrome (ACS) and stable ischemic heart disease.
High-sensitivity assays for cardiac troponin enable more precise measurement of very low concentrations and improved diagnostic accuracy, but the prognostic value of these measurements, particularly at low concentrations, is less well defined.
Scientists at Brigham and Women's Hospital (Boston, MA, USA) evaluated the prognostic performance of a new high-sensitivity cardiac troponin I (hs-cTnI) assay compared with a fourth-generation commercial cardiac-specific isoforms of troponin T (cTnT) assay in 4,695 patients with non-ST-segment elevation acute coronary syndromes (NSTE-ACS) from the Early Glycoprotein IIb/IIIa Inhibition in NSTE-ACS (EARLY-ACS) and the Otamixaban for the Treatment of Patients with NSTE-ACS (SEPIA-ACS1-TIMI) trials. Baseline cardiac troponin was categorized at the 99th percentile reference limit (26 ng/L for hs-cTnI; 10 ng/L for cTnT) and at sex-specific 99th percentiles for hs-cTnI.
The ARCHITECT STAT High Sensitive Troponin-I assay (hs-cTnI, Abbott; Abbott Park, IL, USA) is a two-step, sandwich chemiluminescent magnetic-microparticle immunoassay that uses two mouse monoclonal antibodies recognizing amino acids 41–49 (capture) and 24–40 (detection) on human cTnI, and a limit of detection of 1.5 ng/L. Troponin T was also measured, using the cTnT, fourth generation assay (Roche Diagnostics; Indianapolis, IN, USA), in 4,221 of the 4,695 study participants with hs-cTnI data; and this assay has a limit of detection of 10 ng/L.
All patients at baseline had detectable hs-cTnI compared with 94.5% with detectable cTnT. At 30 days, patients with a baseline hs-cTnI equal to or greater than 26 ng/L had a significantly higher rate of cardiovascular death or MI compared with those with a baseline measurement below the reference limit. The authors concluded that the assay performed well prognostically, identifying patients at significantly higher 30-day risk of cardiovascular death or new MI using the guideline-based 99th percentile decision limit. The results of the study demonstrate a gradient of risk at very low ranges of concentrations, and performs better with respect to prognosis than a current commercial assay. The study was published in the January 2014 issue of the journal Clinical Chemistry.
Related Links:
Brigham and Women's Hospital
Abbott
Roche Diagnostics
High-sensitivity assays for cardiac troponin enable more precise measurement of very low concentrations and improved diagnostic accuracy, but the prognostic value of these measurements, particularly at low concentrations, is less well defined.
Scientists at Brigham and Women's Hospital (Boston, MA, USA) evaluated the prognostic performance of a new high-sensitivity cardiac troponin I (hs-cTnI) assay compared with a fourth-generation commercial cardiac-specific isoforms of troponin T (cTnT) assay in 4,695 patients with non-ST-segment elevation acute coronary syndromes (NSTE-ACS) from the Early Glycoprotein IIb/IIIa Inhibition in NSTE-ACS (EARLY-ACS) and the Otamixaban for the Treatment of Patients with NSTE-ACS (SEPIA-ACS1-TIMI) trials. Baseline cardiac troponin was categorized at the 99th percentile reference limit (26 ng/L for hs-cTnI; 10 ng/L for cTnT) and at sex-specific 99th percentiles for hs-cTnI.
The ARCHITECT STAT High Sensitive Troponin-I assay (hs-cTnI, Abbott; Abbott Park, IL, USA) is a two-step, sandwich chemiluminescent magnetic-microparticle immunoassay that uses two mouse monoclonal antibodies recognizing amino acids 41–49 (capture) and 24–40 (detection) on human cTnI, and a limit of detection of 1.5 ng/L. Troponin T was also measured, using the cTnT, fourth generation assay (Roche Diagnostics; Indianapolis, IN, USA), in 4,221 of the 4,695 study participants with hs-cTnI data; and this assay has a limit of detection of 10 ng/L.
All patients at baseline had detectable hs-cTnI compared with 94.5% with detectable cTnT. At 30 days, patients with a baseline hs-cTnI equal to or greater than 26 ng/L had a significantly higher rate of cardiovascular death or MI compared with those with a baseline measurement below the reference limit. The authors concluded that the assay performed well prognostically, identifying patients at significantly higher 30-day risk of cardiovascular death or new MI using the guideline-based 99th percentile decision limit. The results of the study demonstrate a gradient of risk at very low ranges of concentrations, and performs better with respect to prognosis than a current commercial assay. The study was published in the January 2014 issue of the journal Clinical Chemistry.
Related Links:
Brigham and Women's Hospital
Abbott
Roche Diagnostics
Latest Clinical Chem. News
- 3D Printed Point-Of-Care Mass Spectrometer Outperforms State-Of-The-Art Models
- POC Biomedical Test Spins Water Droplet Using Sound Waves for Cancer Detection
- Highly Reliable Cell-Based Assay Enables Accurate Diagnosis of Endocrine Diseases
- New Blood Testing Method Detects Potent Opioids in Under Three Minutes
- Wireless Hepatitis B Test Kit Completes Screening and Data Collection in One Step
- Pain-Free, Low-Cost, Sensitive, Radiation-Free Device Detects Breast Cancer in Urine
- Spit Test Detects Breast Cancer in Five Seconds
- Electrochemical Sensors with Next-Generation Coating Advances Precision Diagnostics at POC
- First-Of-Its-Kind Handheld Device Accurately Detects Fentanyl in Urine within Seconds
- New Fluorescent Sensor Array Lights up Alzheimer’s-Related Proteins for Earlier Detection
- Automated Mass Spectrometry-Based Clinical Analyzer Could Transform Lab Testing
- Highly Sensitive pH Sensor to Aid Detection of Cancers and Vector-Borne Viruses
- Non-Invasive Sensor Monitors Changes in Saliva Compositions to Rapidly Diagnose Diabetes
- Breakthrough Immunoassays to Aid in Risk Assessment of Preeclampsia
- Urine Test for Monitoring Changes in Kidney Health Markers Can Predict New-Onset Heart Failure
- AACC Releases Comprehensive Diabetes Testing Guidelines