Point-of-Care Device Compared for Cardiac Troponin Assay
By LabMedica International staff writers Posted on 02 Oct 2013 |
Image: The i-STAT portable clinical analyzer system (Photo courtesy of Abbot Laboratories).
A fingerstick point-of-care (POC) testing for cardiac troponin I has been compared to conventional venipuncture POC testing.
Cardiac troponin assays commonly exceed one hour while point-of-care testing can be completed in minutes by basic life support personnel so when patients present complaining of chest pain, time can be critical to achieving positive outcomes.
Scientists at the Loyola University Medical Center (Maywood, IL, USA) collected fingerstick blood samples from consenting patients for whom standard-of-care venipuncture POC troponin (POCT) testing had been ordered as part of their workup. Cardiac troponin (cTnl) assays were performed using a commercial device from June to August 2011. Eighty-nine cTnI levels were measured by both fingerstick and standard venipuncture emergency department POC testing. Four resulted in cartridge error so only the remaining 85 were analyzed.
Cardiac troponin I (cTnI) assays were performed using an i-STAT 1 device (Abbott Point of Care; Princeton, NJ, USA). Fingerstick testing, compared with standard ED POCT, has a positive predictive value of 1.00 (0.48, 1.00), a negative predictive value of 0.96 (0.89, 0.99), a sensitivity of 0.625 (0.24, 0.91), and a specificity of 1.00 (0.95, 1.00). The relationship between methods appeared to be linear.
The authors concluded that fingerstick cTnI testing using the i-STAT device is not accurate enough to determine the exact troponin level without the application of a corrective term. Fingerstick testing is, however, accurate in qualifying troponin levels as negative, borderline, or positive and is, therefore, capable of providing clinical information that may guide diagnostic and therapeutic decision-making. The study was published in the August 2013 issue of the American Journal of Emergency Medicine.
Related Links:
Loyola University Medical Center
Abbott Point of Care
Cardiac troponin assays commonly exceed one hour while point-of-care testing can be completed in minutes by basic life support personnel so when patients present complaining of chest pain, time can be critical to achieving positive outcomes.
Scientists at the Loyola University Medical Center (Maywood, IL, USA) collected fingerstick blood samples from consenting patients for whom standard-of-care venipuncture POC troponin (POCT) testing had been ordered as part of their workup. Cardiac troponin (cTnl) assays were performed using a commercial device from June to August 2011. Eighty-nine cTnI levels were measured by both fingerstick and standard venipuncture emergency department POC testing. Four resulted in cartridge error so only the remaining 85 were analyzed.
Cardiac troponin I (cTnI) assays were performed using an i-STAT 1 device (Abbott Point of Care; Princeton, NJ, USA). Fingerstick testing, compared with standard ED POCT, has a positive predictive value of 1.00 (0.48, 1.00), a negative predictive value of 0.96 (0.89, 0.99), a sensitivity of 0.625 (0.24, 0.91), and a specificity of 1.00 (0.95, 1.00). The relationship between methods appeared to be linear.
The authors concluded that fingerstick cTnI testing using the i-STAT device is not accurate enough to determine the exact troponin level without the application of a corrective term. Fingerstick testing is, however, accurate in qualifying troponin levels as negative, borderline, or positive and is, therefore, capable of providing clinical information that may guide diagnostic and therapeutic decision-making. The study was published in the August 2013 issue of the American Journal of Emergency Medicine.
Related Links:
Loyola University Medical Center
Abbott Point of Care
Latest Clinical Chem. News
- 3D Printed Point-Of-Care Mass Spectrometer Outperforms State-Of-The-Art Models
- POC Biomedical Test Spins Water Droplet Using Sound Waves for Cancer Detection
- Highly Reliable Cell-Based Assay Enables Accurate Diagnosis of Endocrine Diseases
- New Blood Testing Method Detects Potent Opioids in Under Three Minutes
- Wireless Hepatitis B Test Kit Completes Screening and Data Collection in One Step
- Pain-Free, Low-Cost, Sensitive, Radiation-Free Device Detects Breast Cancer in Urine
- Spit Test Detects Breast Cancer in Five Seconds
- Electrochemical Sensors with Next-Generation Coating Advances Precision Diagnostics at POC
- First-Of-Its-Kind Handheld Device Accurately Detects Fentanyl in Urine within Seconds
- New Fluorescent Sensor Array Lights up Alzheimer’s-Related Proteins for Earlier Detection
- Automated Mass Spectrometry-Based Clinical Analyzer Could Transform Lab Testing
- Highly Sensitive pH Sensor to Aid Detection of Cancers and Vector-Borne Viruses
- Non-Invasive Sensor Monitors Changes in Saliva Compositions to Rapidly Diagnose Diabetes
- Breakthrough Immunoassays to Aid in Risk Assessment of Preeclampsia
- Urine Test for Monitoring Changes in Kidney Health Markers Can Predict New-Onset Heart Failure
- AACC Releases Comprehensive Diabetes Testing Guidelines