Enhanced Immunoassay Validated for NGAL
By LabMedica International staff writers Posted on 28 Jun 2017 |
Image: The Gentian NGAL Immunoassay is a new particle-enhanced turbidimetric immunoassay (PETIA) for in vitro diagnostic testing of neutrophil gelatinase-associated lipocalin (NGAL) in human urine and plasma samples (Photo courtesy of Gentian).
The ectopic expression of Neutrophil Gelatinase-Associated Lipocalin (NGAL) has been reported in patients with various malignancies such as kidney, colorectal, gastric, pancreatobiliary, hepatic, breast, gynecological and hematologic cancers.
The concentration of the biomarker NGAL increases considerably immediately after acute kidney injury (AKI). The measurement of NGAL in serum or plasma is also considered to be more convenient than measurement in urine, since urine output may be considerably reduced in patients with AKI.
Scientists at the University of Verona (Italy) designed a study to validate the analytical performance of the new particle-enhanced enhanced turbidimetric immunoassay (PETIA) for measuring neutrophil gelatinase-associated lipocalin (NGAL) in serum samples. The analytical validation of the NGAL PETIA consisted of calculation of limit of blank (LOB), limit of detection (LOD) and functional sensitivity, and estimation of inter-assay and intra-assay imprecision, linearity, and correlation with another widely used NGAL PETIA.
The Gentian NGAL is a particle-enhanced turbidimetric immunoassay (PETIA), which can be applied to a vast array of automated clinical chemistry analyzers. Analytical validation of the Gentian NGAL assay was carried out on a Cobas c501. The quoted NGAL serum/plasma reference range is 13.7 ng/mL to 104.6 ng/mL, the total sample volume needed for the assay is 3 μL and test results are available in 10 minutes.
An excellent correlation was observed between values measured with Gentian NGAL and BioPorto NGAL in 74 routine serum samples. The LOB and LOD of Gentian NGAL were found to be 3.8 ng/mL and 6.3 ng/mL, respectively. An analytical coefficient of variation (CV) of 20% corresponded to a NGAL value of 10 ng/mL. The intra-assay and inter-assay imprecision (CV) was between 0.4% and 5.2% and 0.6% and 7.1% and the total imprecision (CV) was 3.7%. The linearity was optimal at NGAL concentrations between 37 and 1,420 ng/mL.
The authors concluded that the result of their analytical validation show that Gentian NGAL may be a viable option for both routine and urgent assessment of serum NGAL. Additional advantages are the very fast turnaround time as results are available in approximately 10 minutes compared to 35 minutes in another commercial assay and two hours in an enzyme-linked immunosorbent assay (ELISA). The possibility of widespread implementation of this test on a wide array of clinical chemistry platforms may also permit the concomitant measurement of NGAL and other biomarkers of renal function. The full study will be published in the August 2017 issue of the journal Practical Laboratory Medicine.
Related Links:
University of Verona
The concentration of the biomarker NGAL increases considerably immediately after acute kidney injury (AKI). The measurement of NGAL in serum or plasma is also considered to be more convenient than measurement in urine, since urine output may be considerably reduced in patients with AKI.
Scientists at the University of Verona (Italy) designed a study to validate the analytical performance of the new particle-enhanced enhanced turbidimetric immunoassay (PETIA) for measuring neutrophil gelatinase-associated lipocalin (NGAL) in serum samples. The analytical validation of the NGAL PETIA consisted of calculation of limit of blank (LOB), limit of detection (LOD) and functional sensitivity, and estimation of inter-assay and intra-assay imprecision, linearity, and correlation with another widely used NGAL PETIA.
The Gentian NGAL is a particle-enhanced turbidimetric immunoassay (PETIA), which can be applied to a vast array of automated clinical chemistry analyzers. Analytical validation of the Gentian NGAL assay was carried out on a Cobas c501. The quoted NGAL serum/plasma reference range is 13.7 ng/mL to 104.6 ng/mL, the total sample volume needed for the assay is 3 μL and test results are available in 10 minutes.
An excellent correlation was observed between values measured with Gentian NGAL and BioPorto NGAL in 74 routine serum samples. The LOB and LOD of Gentian NGAL were found to be 3.8 ng/mL and 6.3 ng/mL, respectively. An analytical coefficient of variation (CV) of 20% corresponded to a NGAL value of 10 ng/mL. The intra-assay and inter-assay imprecision (CV) was between 0.4% and 5.2% and 0.6% and 7.1% and the total imprecision (CV) was 3.7%. The linearity was optimal at NGAL concentrations between 37 and 1,420 ng/mL.
The authors concluded that the result of their analytical validation show that Gentian NGAL may be a viable option for both routine and urgent assessment of serum NGAL. Additional advantages are the very fast turnaround time as results are available in approximately 10 minutes compared to 35 minutes in another commercial assay and two hours in an enzyme-linked immunosorbent assay (ELISA). The possibility of widespread implementation of this test on a wide array of clinical chemistry platforms may also permit the concomitant measurement of NGAL and other biomarkers of renal function. The full study will be published in the August 2017 issue of the journal Practical Laboratory Medicine.
Related Links:
University of Verona
Latest Clinical Chem. News
- 3D Printed Point-Of-Care Mass Spectrometer Outperforms State-Of-The-Art Models
- POC Biomedical Test Spins Water Droplet Using Sound Waves for Cancer Detection
- Highly Reliable Cell-Based Assay Enables Accurate Diagnosis of Endocrine Diseases
- New Blood Testing Method Detects Potent Opioids in Under Three Minutes
- Wireless Hepatitis B Test Kit Completes Screening and Data Collection in One Step
- Pain-Free, Low-Cost, Sensitive, Radiation-Free Device Detects Breast Cancer in Urine
- Spit Test Detects Breast Cancer in Five Seconds
- Electrochemical Sensors with Next-Generation Coating Advances Precision Diagnostics at POC
- First-Of-Its-Kind Handheld Device Accurately Detects Fentanyl in Urine within Seconds
- New Fluorescent Sensor Array Lights up Alzheimer’s-Related Proteins for Earlier Detection
- Automated Mass Spectrometry-Based Clinical Analyzer Could Transform Lab Testing
- Highly Sensitive pH Sensor to Aid Detection of Cancers and Vector-Borne Viruses
- Non-Invasive Sensor Monitors Changes in Saliva Compositions to Rapidly Diagnose Diabetes
- Breakthrough Immunoassays to Aid in Risk Assessment of Preeclampsia
- Urine Test for Monitoring Changes in Kidney Health Markers Can Predict New-Onset Heart Failure
- AACC Releases Comprehensive Diabetes Testing Guidelines