Serum Prostate-Specific Antigen Measured by Tandem Mass Spectrometry
By LabMedica International staff writers Posted on 23 Oct 2014 |
Prostate-specific antigen (PSA) is a widely used blood test for detection and monitoring of prostate disease and many clinicians assume that all test methods produce essentially the same results, though differences do exist.
PSA is a 34-kDa glycoprotein with chymotrypsin-like enzyme activity that circulates both in free forms and complexed to various enzyme inhibitors including anti-chymotrypsin and α2-macroglobulin. Prostate-specific antigen bound to α2-macroglobulin is not detected by commercial PSA immunoassays.
Scientists at the Mayo Clinic College of Medicine (Rochester, MN, USA) developed a strategy to use a mass spectrometry (MS) assay to measure PSA concentrations that measures only the immune-reactive forms. Prostate-specific antigen is immune extracted from serum using antibodies directed to different PSA epitopes. A combination of three monoclonal anti-PSA antisera, which are directed to different PSA epitopes, was used for immune extraction of PSA from serum. PSA was extracted from serum, trypsin digested, and the LSEPAELTDAVK peptide was quantitated.
Absolute quantitation was performed by liquid chromatography–tandem mass spectrometry using a CTC Analytics HTC PAL auto sampler (LEAP Technologies; Carrboro, NC, USA; www.leaptec.com), a Shimadzu 10-AD binary pumping system (Shimadzu Scientific Instruments; Columbia, MD, USA), and an API 5000 triple quadruple mass spectrometer (AB SCIEX; Framingham, MA, USA). Validation of clinical utility and comparisons with two immunoassays were performed using frozen sera aliquots from 100 men undergoing prostate biopsy, 50 negative, 50 with cancer, and five serial samples collected over time from five men with advanced prostate cancer.
The antibody extraction efficiency was greater than 99%. The assay has an analytic range from 1.2 to 76 ng/mL, with precision ranging from 8.6% at 1.5 ng/mL to 5.4% at 27 ng/mL. The mass spectrometry assay correlated well with the two commercial immunoassays. All three assays showed statistically equivalent separation of prostate cancer from benign disease using receiver operating characteristic curve analysis.
The authors concluded that their MS assays are the first potential reference mass spectrometry assays that measure the same forms of PSA as the commercial immunoassays. These assays can reliably quantitate PSA concentrations in male sera that closely match existing immunoassays, and these assays are traceable to the international PSA standard with the proteins and therefore make the measurement system incomplete. Another advantage of their immune extraction is the increase in the relative concentration of PSA caused when the peptides from the PSA in a larger sample of serum are re-suspended in a smaller injection volume of buffer for the MS. This increase in concentration helps improve the robustness of the MS assay. The study was published on the October 2014 issue of the journal Archives of Pathology & Laboratory Medicine.
Related Links:
Mayo Clinic College of Medicine
LEAP Technologies
AB SCIEX
PSA is a 34-kDa glycoprotein with chymotrypsin-like enzyme activity that circulates both in free forms and complexed to various enzyme inhibitors including anti-chymotrypsin and α2-macroglobulin. Prostate-specific antigen bound to α2-macroglobulin is not detected by commercial PSA immunoassays.
Scientists at the Mayo Clinic College of Medicine (Rochester, MN, USA) developed a strategy to use a mass spectrometry (MS) assay to measure PSA concentrations that measures only the immune-reactive forms. Prostate-specific antigen is immune extracted from serum using antibodies directed to different PSA epitopes. A combination of three monoclonal anti-PSA antisera, which are directed to different PSA epitopes, was used for immune extraction of PSA from serum. PSA was extracted from serum, trypsin digested, and the LSEPAELTDAVK peptide was quantitated.
Absolute quantitation was performed by liquid chromatography–tandem mass spectrometry using a CTC Analytics HTC PAL auto sampler (LEAP Technologies; Carrboro, NC, USA; www.leaptec.com), a Shimadzu 10-AD binary pumping system (Shimadzu Scientific Instruments; Columbia, MD, USA), and an API 5000 triple quadruple mass spectrometer (AB SCIEX; Framingham, MA, USA). Validation of clinical utility and comparisons with two immunoassays were performed using frozen sera aliquots from 100 men undergoing prostate biopsy, 50 negative, 50 with cancer, and five serial samples collected over time from five men with advanced prostate cancer.
The antibody extraction efficiency was greater than 99%. The assay has an analytic range from 1.2 to 76 ng/mL, with precision ranging from 8.6% at 1.5 ng/mL to 5.4% at 27 ng/mL. The mass spectrometry assay correlated well with the two commercial immunoassays. All three assays showed statistically equivalent separation of prostate cancer from benign disease using receiver operating characteristic curve analysis.
The authors concluded that their MS assays are the first potential reference mass spectrometry assays that measure the same forms of PSA as the commercial immunoassays. These assays can reliably quantitate PSA concentrations in male sera that closely match existing immunoassays, and these assays are traceable to the international PSA standard with the proteins and therefore make the measurement system incomplete. Another advantage of their immune extraction is the increase in the relative concentration of PSA caused when the peptides from the PSA in a larger sample of serum are re-suspended in a smaller injection volume of buffer for the MS. This increase in concentration helps improve the robustness of the MS assay. The study was published on the October 2014 issue of the journal Archives of Pathology & Laboratory Medicine.
Related Links:
Mayo Clinic College of Medicine
LEAP Technologies
AB SCIEX
Read the full article by registering today, it's FREE!
Register now for FREE to LabMedica.com and get complete access to news and events that shape the world of Clinical Laboratory Medicine.
- Free digital version edition of LabMedica International sent by email on regular basis
- Free print version of LabMedica International magazine (available only outside USA and Canada).
- Free and unlimited access to back issues of LabMedica International in digital format
- Free LabMedica International Newsletter sent every week containing the latest news
- Free breaking news sent via email
- Free access to Events Calendar
- Free access to LinkXpress new product services
- REGISTRATION IS FREE AND EASY!
Sign in: Registered website members
Sign in: Registered magazine subscribers
Latest Clinical Chem. News
- 3D Printed Point-Of-Care Mass Spectrometer Outperforms State-Of-The-Art Models
- POC Biomedical Test Spins Water Droplet Using Sound Waves for Cancer Detection
- Highly Reliable Cell-Based Assay Enables Accurate Diagnosis of Endocrine Diseases
- New Blood Testing Method Detects Potent Opioids in Under Three Minutes
- Wireless Hepatitis B Test Kit Completes Screening and Data Collection in One Step
- Pain-Free, Low-Cost, Sensitive, Radiation-Free Device Detects Breast Cancer in Urine
- Spit Test Detects Breast Cancer in Five Seconds
- Electrochemical Sensors with Next-Generation Coating Advances Precision Diagnostics at POC
- First-Of-Its-Kind Handheld Device Accurately Detects Fentanyl in Urine within Seconds
- New Fluorescent Sensor Array Lights up Alzheimer’s-Related Proteins for Earlier Detection
- Automated Mass Spectrometry-Based Clinical Analyzer Could Transform Lab Testing
- Highly Sensitive pH Sensor to Aid Detection of Cancers and Vector-Borne Viruses
- Non-Invasive Sensor Monitors Changes in Saliva Compositions to Rapidly Diagnose Diabetes
- Breakthrough Immunoassays to Aid in Risk Assessment of Preeclampsia
- Urine Test for Monitoring Changes in Kidney Health Markers Can Predict New-Onset Heart Failure
- AACC Releases Comprehensive Diabetes Testing Guidelines