Automated Methods Evaluated for Von Willebrand Factor Activity
By LabMedica International staff writers Posted on 12 Jun 2014 |
Image: The Sysmex CS2000i fully automated blood coagulation analyzer (Photo courtesy of Siemens).
Von Willebrand factor (VWF) assays have an important role in the diagnostic evaluation, and treatment monitoring, of von Willebrand disease (VWD), which is one of the most common coagulopathies or bleeding disorders.
The laboratory evaluation for VWD requires an assessment of plasma VWF activity, which is commonly done using an aggregometer to assess VWF ristocetin cofactor activity which is a quantitative method that uses the antibiotic ristocetin to induce plasma VWF binding to the VWF receptor on target platelets and platelet agglutination.
Scientists at the McMaster University (Hamilton, ON, Canada) evaluated 262 samples from 217 patients that included 188 samples from 153 females and 73 samples from 63 males. There were 197 samples from adults (ages 18 to 84) and 64 from children (ages 0 to 17 years), and 67 samples were from 34 patients with previously diagnosed VWD. After validating that the assay could be performed on an instrument from a different manufacturer, they compared VWF activity assay (VWF:Ac) to VWF ristocetin cofactor activity (VWF:RCo) findings, including ratios of activity/antigen.
Plasma was tested by the VWF:Ac on an STA-R Evolution (Diagnostica Stago; Parsippany, NJ, USA) and using the Innovance assay on a Sysmex CS2000i instrument (Siemens; Erlangen, Germany). VWF:RCo was performed by aggregometry on a Helena AggRAM instrument (Helena Laboratories, Beaumont, TX, USA). There was excellent correlation between VWF:Ac results run at two different sites on two different instruments. VWF:Ac had greater precision and sensitivity to low levels of VWF than the VWF:RCo method.
Although there was good correlation between VWF:Ac and VWF:RCo results among healthy controls and patient subjects, VWF:Ac results were undetectable and/or significantly lower than VWF:RCo among patients who had types 2A, 2B, or 2M VWD. Additionally, a higher proportion of patient samples were classified as showing qualitative defects using the VWF:Ac compared with VWF:RCo method.
The authors concluded that the Innovance VWF:Ac method is an acceptable, automated alternative to the VWF:RCo method for assessment of VWF binding to the platelet receptor glycoprotein GPIbα that is sensitive to both quantitative and qualitative defects of VWF. Laboratories and clinicians need to be aware that some patients with VWD have much lower levels of VWF measured by the VWF:Ac assay than by VWF:RCo. The study was published on April 18, 2014, in the International Journal of Laboratory Hematology.
Related Links:
McMaster University
Diagnostica Stago
Siemens Healthcare
The laboratory evaluation for VWD requires an assessment of plasma VWF activity, which is commonly done using an aggregometer to assess VWF ristocetin cofactor activity which is a quantitative method that uses the antibiotic ristocetin to induce plasma VWF binding to the VWF receptor on target platelets and platelet agglutination.
Scientists at the McMaster University (Hamilton, ON, Canada) evaluated 262 samples from 217 patients that included 188 samples from 153 females and 73 samples from 63 males. There were 197 samples from adults (ages 18 to 84) and 64 from children (ages 0 to 17 years), and 67 samples were from 34 patients with previously diagnosed VWD. After validating that the assay could be performed on an instrument from a different manufacturer, they compared VWF activity assay (VWF:Ac) to VWF ristocetin cofactor activity (VWF:RCo) findings, including ratios of activity/antigen.
Plasma was tested by the VWF:Ac on an STA-R Evolution (Diagnostica Stago; Parsippany, NJ, USA) and using the Innovance assay on a Sysmex CS2000i instrument (Siemens; Erlangen, Germany). VWF:RCo was performed by aggregometry on a Helena AggRAM instrument (Helena Laboratories, Beaumont, TX, USA). There was excellent correlation between VWF:Ac results run at two different sites on two different instruments. VWF:Ac had greater precision and sensitivity to low levels of VWF than the VWF:RCo method.
Although there was good correlation between VWF:Ac and VWF:RCo results among healthy controls and patient subjects, VWF:Ac results were undetectable and/or significantly lower than VWF:RCo among patients who had types 2A, 2B, or 2M VWD. Additionally, a higher proportion of patient samples were classified as showing qualitative defects using the VWF:Ac compared with VWF:RCo method.
The authors concluded that the Innovance VWF:Ac method is an acceptable, automated alternative to the VWF:RCo method for assessment of VWF binding to the platelet receptor glycoprotein GPIbα that is sensitive to both quantitative and qualitative defects of VWF. Laboratories and clinicians need to be aware that some patients with VWD have much lower levels of VWF measured by the VWF:Ac assay than by VWF:RCo. The study was published on April 18, 2014, in the International Journal of Laboratory Hematology.
Related Links:
McMaster University
Diagnostica Stago
Siemens Healthcare
Latest Hematology News
- Next Generation Instrument Screens for Hemoglobin Disorders in Newborns
- First 4-in-1 Nucleic Acid Test for Arbovirus Screening to Reduce Risk of Transfusion-Transmitted Infections
- POC Finger-Prick Blood Test Determines Risk of Neutropenic Sepsis in Patients Undergoing Chemotherapy
- First Affordable and Rapid Test for Beta Thalassemia Demonstrates 99% Diagnostic Accuracy
- Handheld White Blood Cell Tracker to Enable Rapid Testing For Infections
- Smart Palm-size Optofluidic Hematology Analyzer Enables POCT of Patients’ Blood Cells
- Automated Hematology Platform Offers High Throughput Analytical Performance
- New Tool Analyzes Blood Platelets Faster, Easily and Accurately
- First Rapid-Result Hematology Analyzer Reports Measures of Infection and Severity at POC
- Bleeding Risk Diagnostic Test to Reduce Preventable Complications in Hospitals
- True POC Hematology Analyzer with Direct Capillary Sampling Enhances Ease-of-Use and Testing Throughput
- Point of Care CBC Analyzer with Direct Capillary Sampling Enhances Ease-of-Use and Testing Throughput
- Blood Test Could Predict Outcomes in Emergency Department and Hospital Admissions
- Novel Technology Diagnoses Immunothrombosis Using Breath Gas Analysis
- Advanced Hematology System Allows Labs to Process Up To 119 Complete Blood Count Results per Hour
- Unique AI-Based Approach Automates Clinical Analysis of Blood Data