Newly Developed Point-of-Care Hemoglobinometer Evaluated

By LabMedica International staff writers
Posted on 10 Dec 2014
Print article
Image: The TrueHb hemometer testing kit (Photo courtesy of the Indian Institute of Technology Delhi).
Image: The TrueHb hemometer testing kit (Photo courtesy of the Indian Institute of Technology Delhi).
Hemoglobin (Hb) concentration is routinely measured using automated analyzers, and although these counters are very accurate and reliable, they are expensive and too cumbersome for portability.

Portable hemoglobinometer can potentially provide the solution to the challenges as they are affordable, precise, and convenient and require only a tiny sample of capillary or venous blood, do not require any calibration and display digital results immediately.

Scientists at the Indian Institute of Technology Delhi (IITD; New Delhi, India) collected 200 hundred randomly selected venous blood samples, submitted for hemogram study at the Department of Hematology, All India Institute of Medical Sciences (New Delhi, India). The hemoglobin values of these samples were evenly spread. There were 67, 76, and 57 samples in the ranges of 3.4–8 g/dL, 8–12.5 g/dL, and 12.5–20 g/dL respectively.

The investigators evaluated the performance of the portable hemometer called TrueHb (New Delhi, India) developed by IITD, and compared it with an automated five-part hematology analyzer intended for in vitro diagnostic use in clinical laboratories, the Sysmex counter XT 1800i (Sysmex; Kobe, Japan). The two set of values were comparatively analyzed and the repeatability of the performance of TrueHb was also evaluated against Sysmex value.

The scatter plot of TrueHb values and Sysmex values showed linear distribution with positive correlations. The intraclass correlation (ICC) values between the two set of values was found to be very high. The mean difference in Bland–Altman plots of TrueHb values against the Sysmex values was found to be −0.02, with limits of agreement between -0.777 and 0.732 g/dL. Statistical analysis suggested good repeatability in results of TrueHb, having a low mean coefficient of variation (CV) of 2.22, against 4.44, that of Sysmex values, and 95% confidence interval of 1.99 to 2.44, against 3.85 to 5.03, that of Sysmex values.

The authors concluded that a strong positive correlation between the two measurements devices suggest that the newly developed hemometer, TrueHb, can potentially replace a pathology laboratory analyzer for the purpose of measuring hemoglobin in blood samples. Keeping the performance of TrueHb in view, the newly developed device exhibited the potential to become a dependable tool for healthcare providers for measuring and monitoring hemoglobin levels of patients in conventional healthcare setups as well as in field studies. The study was published on November 22, 2014, in the International Journal of Laboratory Hematology.

Related Links:
Indian Institute of Technology Delhi 
All India Institute of Medical Sciences
TrueHb



Platinum Member
COVID-19 Rapid Test
OSOM COVID-19 Antigen Rapid Test
Magnetic Bead Separation Modules
MAG and HEATMAG
Complement 3 (C3) Test
GPP-100 C3 Kit
Gold Member
Real-time PCR System
GentierX3 Series

Print article

Channels

Clinical Chemistry

view channel
Image: The 3D printed miniature ionizer is a key component of a mass spectrometer (Photo courtesy of MIT)

3D Printed Point-Of-Care Mass Spectrometer Outperforms State-Of-The-Art Models

Mass spectrometry is a precise technique for identifying the chemical components of a sample and has significant potential for monitoring chronic illness health states, such as measuring hormone levels... Read more

Molecular Diagnostics

view channel
Image: A massive study has identified new biomarkers for renal cancer subtypes, improving diagnosis and treatment (Photo courtesy of Jessica Johnson)

Novel Biomarkers to Improve Diagnosis of Renal Cell Carcinoma Subtypes

Renal cell carcinomas (RCCs) are notably diverse, encompassing over 20 distinct subtypes and generally categorized into clear cell and non-clear cell types; around 20% of all RCCs fall into the non-clear... Read more

Immunology

view channel
Image: Exosomes can be a promising biomarker for cellular rejection after organ transplant (Photo courtesy of Nicolas Primola/Shutterstock)

Diagnostic Blood Test for Cellular Rejection after Organ Transplant Could Replace Surgical Biopsies

Transplanted organs constantly face the risk of being rejected by the recipient's immune system which differentiates self from non-self using T cells and B cells. T cells are commonly associated with acute... Read more

Microbiology

view channel
Image: The T-SPOT.TB test is now paired with the Auto-Pure 2400 liquid handling platform for accurate TB testing (Photo courtesy of Shutterstock)

Integrated Solution Ushers New Era of Automated Tuberculosis Testing

Tuberculosis (TB) is responsible for 1.3 million deaths every year, positioning it as one of the top killers globally due to a single infectious agent. In 2022, around 10.6 million people were diagnosed... Read more