Different Test Systems Analytical Performance Evaluated for Serum Creatinine
|
By LabMedica International staff writers Posted on 13 Jan 2022 |

Image: The cobas CREP2 reagent kit: an in vitro test for the quantitative determination of creatinine in serum, human plasma and urine (Photo courtesy of Roche Diagnostics)
Creatinine is a breakdown product of creatine phosphate from muscle and protein metabolism. It is produced and released by the body at a constant rate into the blood, and it is then carried to the kidneys through the blood circulation.
In clinical practice, the level of serum creatinine (SCr) is generally used as a significant indicator for renal function evaluation. SCr determination is a widely applied diagnostic test to evaluate glomerular filtration rate (GFR) that when renal function is impaired, the creatinine level rises. Clinical methods for serum creatinine determination include chemical and enzymatic methods.
Clinical Laboratorians at the University of Chinese Medicine (Guangzhou, China) obtained serum samples by collecting surplus samples from routine creatinine tests. Specimens for method comparison study were collected from left-over clinical 34 patient samples, with creatinine levels ranging from 45 to 1,610 μmol/L. Precision, accuracy, linearity, dilution, lower limit of measurement and analytical interference were studied between two test systems.
The scientists used the open test system, namely the Shino-Test creatinine reagent kit (Shino-Test Corp., Tokyo, Japan), and the Roche Cobas 8000 c702 chemistry analyzer (Roche Diagnostics, Mannheim, Germany) to determine creatinine levels. The determination results were analyzed and compared with those obtained from the matched test system, Roche creatinine reagent kit and the same analyzer. An interference study was carried out by adding different solutions of interferents to serum pools at three levels of creatinine. To study the influence of hemolysis, hemolysate was added to each serum pool aliquot. Interference studies of vitamin C or dobutamine were carried out by mixing serum pool aliquots with serial dilutions of vitamin C solution or dobutamine injection.
The investigators reported that the performance of SCr from the open test system was in compliance with the matched test system with good precision, accuracy, and linearity. In presence of most common interferents, both test systems could lead to accurate creatinine results except for the existence of specified drugs. For dobutamine, the open test system showed better anti-interference performance than the matched system. The results showed that either matched test system or open test system presented extremely low imprecision at each level evaluated.
The authors concluded that their study provided referable opinions for clinical laboratory selection on the test system and a framework for future analogous studies based on different test systems. Clinicians and laboratory professionals should be mindful of potential interference caused by specified medications such as calcium dobesilate and dobutamine because an inaccurate creatinine result may lead to serious untoward consequences. The study was published on December 26, 2021 in the Journal of Clinical Laboratory Analysis.
Related Links:
University of Chinese Medicine
Shino-Test Corp
Roche Diagnostics
In clinical practice, the level of serum creatinine (SCr) is generally used as a significant indicator for renal function evaluation. SCr determination is a widely applied diagnostic test to evaluate glomerular filtration rate (GFR) that when renal function is impaired, the creatinine level rises. Clinical methods for serum creatinine determination include chemical and enzymatic methods.
Clinical Laboratorians at the University of Chinese Medicine (Guangzhou, China) obtained serum samples by collecting surplus samples from routine creatinine tests. Specimens for method comparison study were collected from left-over clinical 34 patient samples, with creatinine levels ranging from 45 to 1,610 μmol/L. Precision, accuracy, linearity, dilution, lower limit of measurement and analytical interference were studied between two test systems.
The scientists used the open test system, namely the Shino-Test creatinine reagent kit (Shino-Test Corp., Tokyo, Japan), and the Roche Cobas 8000 c702 chemistry analyzer (Roche Diagnostics, Mannheim, Germany) to determine creatinine levels. The determination results were analyzed and compared with those obtained from the matched test system, Roche creatinine reagent kit and the same analyzer. An interference study was carried out by adding different solutions of interferents to serum pools at three levels of creatinine. To study the influence of hemolysis, hemolysate was added to each serum pool aliquot. Interference studies of vitamin C or dobutamine were carried out by mixing serum pool aliquots with serial dilutions of vitamin C solution or dobutamine injection.
The investigators reported that the performance of SCr from the open test system was in compliance with the matched test system with good precision, accuracy, and linearity. In presence of most common interferents, both test systems could lead to accurate creatinine results except for the existence of specified drugs. For dobutamine, the open test system showed better anti-interference performance than the matched system. The results showed that either matched test system or open test system presented extremely low imprecision at each level evaluated.
The authors concluded that their study provided referable opinions for clinical laboratory selection on the test system and a framework for future analogous studies based on different test systems. Clinicians and laboratory professionals should be mindful of potential interference caused by specified medications such as calcium dobesilate and dobutamine because an inaccurate creatinine result may lead to serious untoward consequences. The study was published on December 26, 2021 in the Journal of Clinical Laboratory Analysis.
Related Links:
University of Chinese Medicine
Shino-Test Corp
Roche Diagnostics
Latest Clinical Chem. News
- CSF Biomarker Improves Diagnosis of Parkinson’s Disease and Lewy Body Dementia
- Simple Urine Home Test Kit Could Detect Early-Stage Breast Cancer
- New Tool Tracks Biomarker Changes to Predict Myeloma Progression
- New Plasma Tau Assay Improves Prediction of Alzheimer’s Progression
- First IVD Immunoassay to Detect Alzheimer’s Risk Gene Variant Receives CE Mark
- Routine Blood Markers Predict Heart Failure Risk in Prediabetes
- AI Model Enables Personalized Glucose Predictions for Type 1 Diabetes
- AI-Powered Blood Test Distinguishes Deadly Cardiac Events
- AI Sensor Detects Neurological Disorders Using Single Saliva Drop
- Blood Test Tracks Transplant Health Using Donor DNA
- New Blood Test Index Offers Earlier Detection of Liver Scarring
- Electronic Nose Smells Early Signs of Ovarian Cancer in Blood
- Simple Blood Test Offers New Path to Alzheimer’s Assessment in Primary Care
- Existing Hospital Analyzers Can Identify Fake Liquid Medical Products
- Rapid Blood Testing Method Aids Safer Decision-Making in Drug-Related Emergencies
- New PSA-Based Prognostic Model Improves Prostate Cancer Risk Assessment
Channels
Molecular Diagnostics
view channel
New Respiratory Panel Expands Pathogen Detection to 25 Targets
Respiratory infections often present with overlapping symptoms, complicating differential diagnosis in acute and community settings. The stakes are higher for older adults, young children, and people with... Read more
Simple Nasal Swab May Reveal Early Signs of Alzheimer’s Disease
Alzheimer’s disease affects millions worldwide but remains difficult to detect at its earliest, pre-symptomatic stage. Clinicians need tools that can identify biological changes before cognitive symptoms... Read moreHematology
view channel
Rapid Cartridge-Based Test Aims to Expand Access to Hemoglobin Disorder Diagnosis
Sickle cell disease and beta thalassemia are hemoglobin disorders that often require referral to specialized laboratories for definitive diagnosis, delaying results for patients and clinicians.... Read more
New Guidelines Aim to Improve AL Amyloidosis Diagnosis
Light chain (AL) amyloidosis is a rare, life-threatening bone marrow disorder in which abnormal amyloid proteins accumulate in organs. Approximately 3,260 people in the United States are diagnosed... Read moreImmunology
view channel
Study Identifies Inflammatory Pathway Driving Immunotherapy Resistance in Bladder Cancer
Bladder cancer remains a prevalent malignancy with variable responses to immune checkpoint inhibitors. Clinicians often observe elevated C-reactive protein and interleukin-6 in affected patients, yet the... Read more
Microfluidic Chip Detects Cancer Recurrence from Immune Response Signals
Early identification of treatment response and relapse remains a major challenge in solid tumors, where minimal residual disease is difficult to detect with routine imaging and blood tests.... Read moreMicrobiology
view channel
Breath Analysis Approach Offers Rapid Detection of Bacterial Infection
Accurate and rapid identification of bacterial infections remains challenging in acute care, where delays can hinder timely, targeted therapy. Infectious diseases are a major cause of mortality worldwide,... Read more
Study Highlights Accuracy Gaps in Consumer Gut Microbiome Kits
Direct-to-consumer gut microbiome kits promise personalized insights by profiling fecal bacteria and generating health readouts, but their analytical accuracy remains uncertain. A new study shows that... Read more
WHO Recommends Near POC Tests, Tongue Swabs and Sputum Pooling for TB Diagnosis
Tuberculosis (TB) remains one of the world’s leading infectious disease killers, yet millions of cases go undiagnosed or are detected too late. Barriers such as reliance on sputum samples, limited laboratory... Read morePathology
view channel
Biopsy-Based Gene Test Predicts Recurrence Risk in Lung Adenocarcinoma
Lung cancer is the leading cause of cancer death, killing more people in the United States than breast, prostate, and colon cancers combined. In lung adenocarcinoma (LUAD), tumors that invade nearby blood... Read more
AI-Powered Tool to Transform Dermatopathology Workflow
Skin cancer accounts for the largest number of cancer diagnoses in the United States, placing sustained pressure on pathology services. Diagnostic interpretation can be variable for challenging melanocytic... Read moreTechnology
view channel
Portable Breath Sensor Detects Pneumonia Biomarkers in Minutes
Pneumonia is commonly confirmed with chest X-rays or laboratory assays that can take hours, delaying clinical decisions in acute and outpatient settings. Breath-based diagnostics promise faster answers... Read more
New Electronic Pipette Enhances Workflows with Touchscreen Control
Manual pipetting remains a routine yet error-prone step that can affect reproducibility and throughput in clinical and research laboratories. Training demands and ergonomic strain also add variability... Read more
AI Model Outperforms Clinicians in Rare Disease Detection
Rare diseases affect an estimated 300 million people worldwide, yet diagnosis is often protracted and error-prone. Many conditions present with heterogeneous signs that overlap with common disorders, leading... Read more
AI-Driven Diagnostic Demonstrates High Accuracy in Detecting Periprosthetic Joint Infection
Periprosthetic joint infection (PJI) is a rare but serious complication affecting 1% to 2% of primary joint replacement surgeries. The condition occurs when bacteria or fungi infect tissues around an implanted... Read moreIndustry
view channel
Integrated DNA Technologies Expands into Clinical Diagnostics
Integrated DNA Technologies (IDT; Coralville, Iowa, USA) has announced the launch of Archer FUSIONPlex-HT Dx and VARIANTPlex-HT Dx. This launch marks the company’s first in vitro diagnostic (IVD) offerings... Read more