Different Test Systems Analytical Performance Evaluated for Serum Creatinine
|
By LabMedica International staff writers Posted on 13 Jan 2022 |

Image: The cobas CREP2 reagent kit: an in vitro test for the quantitative determination of creatinine in serum, human plasma and urine (Photo courtesy of Roche Diagnostics)
Creatinine is a breakdown product of creatine phosphate from muscle and protein metabolism. It is produced and released by the body at a constant rate into the blood, and it is then carried to the kidneys through the blood circulation.
In clinical practice, the level of serum creatinine (SCr) is generally used as a significant indicator for renal function evaluation. SCr determination is a widely applied diagnostic test to evaluate glomerular filtration rate (GFR) that when renal function is impaired, the creatinine level rises. Clinical methods for serum creatinine determination include chemical and enzymatic methods.
Clinical Laboratorians at the University of Chinese Medicine (Guangzhou, China) obtained serum samples by collecting surplus samples from routine creatinine tests. Specimens for method comparison study were collected from left-over clinical 34 patient samples, with creatinine levels ranging from 45 to 1,610 μmol/L. Precision, accuracy, linearity, dilution, lower limit of measurement and analytical interference were studied between two test systems.
The scientists used the open test system, namely the Shino-Test creatinine reagent kit (Shino-Test Corp., Tokyo, Japan), and the Roche Cobas 8000 c702 chemistry analyzer (Roche Diagnostics, Mannheim, Germany) to determine creatinine levels. The determination results were analyzed and compared with those obtained from the matched test system, Roche creatinine reagent kit and the same analyzer. An interference study was carried out by adding different solutions of interferents to serum pools at three levels of creatinine. To study the influence of hemolysis, hemolysate was added to each serum pool aliquot. Interference studies of vitamin C or dobutamine were carried out by mixing serum pool aliquots with serial dilutions of vitamin C solution or dobutamine injection.
The investigators reported that the performance of SCr from the open test system was in compliance with the matched test system with good precision, accuracy, and linearity. In presence of most common interferents, both test systems could lead to accurate creatinine results except for the existence of specified drugs. For dobutamine, the open test system showed better anti-interference performance than the matched system. The results showed that either matched test system or open test system presented extremely low imprecision at each level evaluated.
The authors concluded that their study provided referable opinions for clinical laboratory selection on the test system and a framework for future analogous studies based on different test systems. Clinicians and laboratory professionals should be mindful of potential interference caused by specified medications such as calcium dobesilate and dobutamine because an inaccurate creatinine result may lead to serious untoward consequences. The study was published on December 26, 2021 in the Journal of Clinical Laboratory Analysis.
Related Links:
University of Chinese Medicine
Shino-Test Corp
Roche Diagnostics
In clinical practice, the level of serum creatinine (SCr) is generally used as a significant indicator for renal function evaluation. SCr determination is a widely applied diagnostic test to evaluate glomerular filtration rate (GFR) that when renal function is impaired, the creatinine level rises. Clinical methods for serum creatinine determination include chemical and enzymatic methods.
Clinical Laboratorians at the University of Chinese Medicine (Guangzhou, China) obtained serum samples by collecting surplus samples from routine creatinine tests. Specimens for method comparison study were collected from left-over clinical 34 patient samples, with creatinine levels ranging from 45 to 1,610 μmol/L. Precision, accuracy, linearity, dilution, lower limit of measurement and analytical interference were studied between two test systems.
The scientists used the open test system, namely the Shino-Test creatinine reagent kit (Shino-Test Corp., Tokyo, Japan), and the Roche Cobas 8000 c702 chemistry analyzer (Roche Diagnostics, Mannheim, Germany) to determine creatinine levels. The determination results were analyzed and compared with those obtained from the matched test system, Roche creatinine reagent kit and the same analyzer. An interference study was carried out by adding different solutions of interferents to serum pools at three levels of creatinine. To study the influence of hemolysis, hemolysate was added to each serum pool aliquot. Interference studies of vitamin C or dobutamine were carried out by mixing serum pool aliquots with serial dilutions of vitamin C solution or dobutamine injection.
The investigators reported that the performance of SCr from the open test system was in compliance with the matched test system with good precision, accuracy, and linearity. In presence of most common interferents, both test systems could lead to accurate creatinine results except for the existence of specified drugs. For dobutamine, the open test system showed better anti-interference performance than the matched system. The results showed that either matched test system or open test system presented extremely low imprecision at each level evaluated.
The authors concluded that their study provided referable opinions for clinical laboratory selection on the test system and a framework for future analogous studies based on different test systems. Clinicians and laboratory professionals should be mindful of potential interference caused by specified medications such as calcium dobesilate and dobutamine because an inaccurate creatinine result may lead to serious untoward consequences. The study was published on December 26, 2021 in the Journal of Clinical Laboratory Analysis.
Related Links:
University of Chinese Medicine
Shino-Test Corp
Roche Diagnostics
Latest Clinical Chem. News
- Multiprotein Plasma Panel Identifies Remission in Vasculitis
- Rapid Urine Biomarker Panel Supports Better Kidney Transplant Decisions
- Blood Test Measures Amyloid Seeding Activity to Detect Alzheimer’s Disease
- FDA Clears Roche Blood Test for Alzheimer’s Amyloid Pathology Assessment
- Blood Test Markers Could Help Identify Older Adults at Risk of Disability
- Rapid Ferritin Test Enables Iron Deficiency Detection at the Point of Care
- Alzheimer’s Blood Test Becomes First FDA-Cleared Option for Adults as Young as 40
- Routine Urine Protein Testing Identifies Undiagnosed Chronic Kidney Disease
- Simple Whole-Blood Screen Identifies Elevated Bile Acids Linked to Liver Disease
- Longitudinal Heart Stress Assessment Improves Prediction of Cardiovascular Outcomes
- Blood Test Measuring Tumor Proliferation May Guide Treatment Sequencing in Metastatic Melanoma
- Head-to-Head Study Compares Blood Tests for Early Alzheimer’s Disease Detection
- Rapid Breath Sensor Detects SARS-CoV-2 from Exhaled Air
- Mathematical Biomarkers Use Routine Blood Tests to Guide Adaptive Prostate Cancer Therapy
- New Chairside Oral Swab Test Could Guide Biopsy Decisions in Oral Cancer
- Multi-Protein Blood Test Improves Staging of Alzheimer’s Disease
Channels
Molecular Diagnostics
view channel
New Blood RNA Markers Help Advance Precision Medicine for Respiratory Patients
Risk stratification in hospitalized respiratory disease, particularly among older adults with COVID-19, remains challenging despite rich clinical datasets. Blood-based non-coding RNA biomarkers are promising,... Read more
Single Genetic Analysis Identifies Causes of Premature Ovarian Insufficiency
Premature ovarian insufficiency (POI) affects up to 3.5% of women and represents a major cause of infertility. In most cases, the underlying etiology remains unknown, making patient counseling and clinical... Read moreHematology
view channel
New Genetic Findings Reveal Cause of Bone Marrow Failure Syndrome
Inherited bone marrow failure syndromes (IBMFS) impair the bone marrow’s ability to produce sufficient healthy blood cells and are associated with an increased risk of early-onset myelodysplastic syndromes (MDS).... Read more
Ultra-Portable Device Enables Finger-Prick Blood Testing at Home
Patients who need frequent blood tests often face repeated clinic visits that burden services and disrupt care. In the UK, millions of tests are performed each year to diagnose disease, guide therapy,... Read moreImmunology
view channel
Gut “Memory” May Explain Why IBD Flares Return
Inflammatory bowel diseases (IBD), including Crohn’s disease and ulcerative colitis, are characterized by recurring flare-ups that damage the intestinal mucosa and remain difficult to predict.... Read more
AI-Based Antibody Profiling Predicts Strength of COVID-19 Vaccine Response
Vaccine-induced protection can vary widely between individuals, creating uncertainty for clinicians, particularly in patients with immunosuppressive conditions. Demographic and health factors explain only... Read moreMicrobiology
view channel
New Diagnostic Workflow Identifies Bloodstream Pathogens and Antibiotic Response in Hours
Sepsis is a life-threatening complication of infection that affects more than 1.5 million patients annually in the United States and contributes to roughly one in three in-hospital deaths.... Read more
Genetic Marker Identifies Emerging Resistance to Frontline Malaria Drugs
Artemisinin-based combination therapy remains central to controlling Plasmodium falciparum malaria, but emerging drug resistance is complicating treatment and surveillance. In Uganda and across sub-Saharan... Read morePathology
view channel
AI Matches Pathologists in Predicting Breast Cancer Prognosis from Immune Cells
Breast cancer is the most common cancer in Australian women, with more than 20,000 cases each year. Prognosis can be informed by counting tumor-infiltrating lymphocytes (TILs) on routine pathology slides,... Read more
AI-Pathologist Framework Improves Accuracy and Reliability in Cancer Diagnosis
Ensuring reliable cancer diagnosis from digital pathology remains a critical challenge as clinical decisions rely on accurate slide interpretation. While artificial intelligence (AI) has accelerated whole-slide... Read more
New Review Highlights Intelligent Agents as Next Step for Digital Pathology
Digital pathology remains constrained by models that classify single images without mirroring how clinicians interrogate multiple slides, adjust magnification, and synthesize ancillary tests.... Read more
Simple Immunohistochemical Size Ratio Improves Classification of Primary Aldosteronism
Primary aldosteronism is a common but underdiagnosed cause of hypertension, in which excess aldosterone promotes salt retention and raises blood pressure. Identifying the dominant source of hormone overproduction... Read moreTechnology
view channel
Autonomous Robotic System Receives FDA Authorization for Blood Collection
Venipuncture is central to many diagnostic pathways, but routine blood collection can be affected by staffing constraints and procedural variability that influence consistency and patient experience.... Read more
Interoperable Data Platform Standardizes Multi-Cancer Blood Test Results
Proteotype Diagnostics has introduced Alchemi, a proprietary data and clinical workflow platform being developed for Enlighten, the company’s investigational blood-based multi-cancer test.... Read moreIndustry
view channelTumor-Informed MRD Assay Gains Medicare Coverage for Immunotherapy Monitoring
NeoGenomics’ RaDaR ST molecular residual disease (MRD) assay has received expanded coverage from the Centers for Medicare & Medicaid Services’ Molecular Diagnostic Services Program (MolDX) for monitoring... Read more