Different Test Systems Analytical Performance Evaluated for Serum Creatinine
|
By LabMedica International staff writers Posted on 13 Jan 2022 |

Image: The cobas CREP2 reagent kit: an in vitro test for the quantitative determination of creatinine in serum, human plasma and urine (Photo courtesy of Roche Diagnostics)
Creatinine is a breakdown product of creatine phosphate from muscle and protein metabolism. It is produced and released by the body at a constant rate into the blood, and it is then carried to the kidneys through the blood circulation.
In clinical practice, the level of serum creatinine (SCr) is generally used as a significant indicator for renal function evaluation. SCr determination is a widely applied diagnostic test to evaluate glomerular filtration rate (GFR) that when renal function is impaired, the creatinine level rises. Clinical methods for serum creatinine determination include chemical and enzymatic methods.
Clinical Laboratorians at the University of Chinese Medicine (Guangzhou, China) obtained serum samples by collecting surplus samples from routine creatinine tests. Specimens for method comparison study were collected from left-over clinical 34 patient samples, with creatinine levels ranging from 45 to 1,610 μmol/L. Precision, accuracy, linearity, dilution, lower limit of measurement and analytical interference were studied between two test systems.
The scientists used the open test system, namely the Shino-Test creatinine reagent kit (Shino-Test Corp., Tokyo, Japan), and the Roche Cobas 8000 c702 chemistry analyzer (Roche Diagnostics, Mannheim, Germany) to determine creatinine levels. The determination results were analyzed and compared with those obtained from the matched test system, Roche creatinine reagent kit and the same analyzer. An interference study was carried out by adding different solutions of interferents to serum pools at three levels of creatinine. To study the influence of hemolysis, hemolysate was added to each serum pool aliquot. Interference studies of vitamin C or dobutamine were carried out by mixing serum pool aliquots with serial dilutions of vitamin C solution or dobutamine injection.
The investigators reported that the performance of SCr from the open test system was in compliance with the matched test system with good precision, accuracy, and linearity. In presence of most common interferents, both test systems could lead to accurate creatinine results except for the existence of specified drugs. For dobutamine, the open test system showed better anti-interference performance than the matched system. The results showed that either matched test system or open test system presented extremely low imprecision at each level evaluated.
The authors concluded that their study provided referable opinions for clinical laboratory selection on the test system and a framework for future analogous studies based on different test systems. Clinicians and laboratory professionals should be mindful of potential interference caused by specified medications such as calcium dobesilate and dobutamine because an inaccurate creatinine result may lead to serious untoward consequences. The study was published on December 26, 2021 in the Journal of Clinical Laboratory Analysis.
Related Links:
University of Chinese Medicine
Shino-Test Corp
Roche Diagnostics
In clinical practice, the level of serum creatinine (SCr) is generally used as a significant indicator for renal function evaluation. SCr determination is a widely applied diagnostic test to evaluate glomerular filtration rate (GFR) that when renal function is impaired, the creatinine level rises. Clinical methods for serum creatinine determination include chemical and enzymatic methods.
Clinical Laboratorians at the University of Chinese Medicine (Guangzhou, China) obtained serum samples by collecting surplus samples from routine creatinine tests. Specimens for method comparison study were collected from left-over clinical 34 patient samples, with creatinine levels ranging from 45 to 1,610 μmol/L. Precision, accuracy, linearity, dilution, lower limit of measurement and analytical interference were studied between two test systems.
The scientists used the open test system, namely the Shino-Test creatinine reagent kit (Shino-Test Corp., Tokyo, Japan), and the Roche Cobas 8000 c702 chemistry analyzer (Roche Diagnostics, Mannheim, Germany) to determine creatinine levels. The determination results were analyzed and compared with those obtained from the matched test system, Roche creatinine reagent kit and the same analyzer. An interference study was carried out by adding different solutions of interferents to serum pools at three levels of creatinine. To study the influence of hemolysis, hemolysate was added to each serum pool aliquot. Interference studies of vitamin C or dobutamine were carried out by mixing serum pool aliquots with serial dilutions of vitamin C solution or dobutamine injection.
The investigators reported that the performance of SCr from the open test system was in compliance with the matched test system with good precision, accuracy, and linearity. In presence of most common interferents, both test systems could lead to accurate creatinine results except for the existence of specified drugs. For dobutamine, the open test system showed better anti-interference performance than the matched system. The results showed that either matched test system or open test system presented extremely low imprecision at each level evaluated.
The authors concluded that their study provided referable opinions for clinical laboratory selection on the test system and a framework for future analogous studies based on different test systems. Clinicians and laboratory professionals should be mindful of potential interference caused by specified medications such as calcium dobesilate and dobutamine because an inaccurate creatinine result may lead to serious untoward consequences. The study was published on December 26, 2021 in the Journal of Clinical Laboratory Analysis.
Related Links:
University of Chinese Medicine
Shino-Test Corp
Roche Diagnostics
Latest Clinical Chem. News
- VOCs Show Promise for Early Multi-Cancer Detection
- Portable Raman Spectroscopy Offers Cost-Effective Kidney Disease Diagnosis at POC
- Gold Nanoparticles to Improve Accuracy of Ovarian Cancer Diagnosis
- Simultaneous Cell Isolation Technology Improves Cancer Diagnostic Accuracy
- Simple Non-Invasive Hair-Based Test Could Speed ALS Diagnosis
- Paper Strip Saliva Test Detects Elevated Uric Acid Levels Without Blood Draws
- Prostate Cancer Markers Based on Chemical Make-Up of Calcifications to Speed Up Detection
- Breath Test Could Help Detect Blood Cancers
- ML-Powered Gas Sensors to Detect Pathogens and AMR at POC
- Saliva-Based Cancer Detection Technology Eliminates Need for Complex Sample Preparation
- Skin Swabs Could Detect Parkinson’s Years Before Symptoms Appear
- New Clinical Chemistry Analyzer Designed to Meet Growing Demands of Modern Labs

- New Reference Measurement Procedure Standardizes Nucleic Acid Amplification Test Results
- Pen-Like Tool Quickly and Non-Invasively Detects Opioids from Skin
- Simple Urine Test Could Detect Multiple Cancers at Early Stage
- Earwax Test Accurately Detects Parkinson’s by Identifying Odor Molecules
Channels
Molecular Diagnostics
view channel
First Direct Measurement of Dementia-Linked Proteins to Enable Early Alzheimer’s Detection
The disease process in Alzheimer’s begins long before memory loss or cognitive decline becomes apparent. During this silent phase, misfolded proteins gradually form amyloid fibrils, which accumulate in... Read more
New Diagnostic Method Detects Pneumonia at POC in Low-Resource Settings
Pneumonia continues to be one of the leading causes of death in low- and middle-income countries, where limited access to advanced laboratory infrastructure hampers early and accurate diagnosis.... Read moreHematology
view channel
ADLM’s New Coagulation Testing Guidance to Improve Care for Patients on Blood Thinners
Direct oral anticoagulants (DOACs) are one of the most common types of blood thinners. Patients take them to prevent a host of complications that could arise from blood clotting, including stroke, deep... Read more
Viscoelastic Testing Could Improve Treatment of Maternal Hemorrhage
Postpartum hemorrhage, severe bleeding after childbirth, remains one of the leading causes of maternal mortality worldwide, yet many of these deaths are preventable. Standard care can be hindered by delays... Read more
Pioneering Model Measures Radiation Exposure in Blood for Precise Cancer Treatments
Scientists have long focused on protecting organs near tumors during radiotherapy, but blood — a vital, circulating tissue — has largely been excluded from dose calculations. Each blood cell passing through... Read moreImmunology
view channel
Chip Captures Cancer Cells from Blood to Help Select Right Breast Cancer Treatment
Ductal carcinoma in situ (DCIS) accounts for about a quarter of all breast cancer cases and generally carries a good prognosis. This non-invasive form of the disease may or may not become life-threatening.... Read more
Blood-Based Liquid Biopsy Model Analyzes Immunotherapy Effectiveness
Immunotherapy has revolutionized cancer care by harnessing the immune system to fight tumors, yet predicting who will benefit remains a major challenge. Many patients undergo costly and taxing treatment... Read moreMicrobiology
view channel
High-Throughput Enteric Panels Detect Multiple GI Bacterial Infections from Single Stool Swab Sample
Gastrointestinal (GI) infections are among the most common causes of illness worldwide, leading to over 1.7 million deaths annually and placing a heavy burden on healthcare systems. Conventional diagnostic... Read more
Fast Noninvasive Bedside Test Uses Sugar Fingerprint to Detect Fungal Infections
Candida bloodstream infections are a growing global health threat, causing an estimated 6 million cases and 3.8 million deaths annually. Hospitals are particularly vulnerable, as weakened patients after... Read morePathology
view channel
New Molecular Analysis Tool to Improve Disease Diagnosis
Accurately distinguishing between similar biomolecules such as proteins is vital for biomedical research and diagnostics, yet existing analytical tools often fail to detect subtle structural or compositional... Read more
Tears Offer Noninvasive Alternative for Diagnosing Neurodegenerative Diseases
Diagnosing and monitoring eye and neurodegenerative diseases often requires invasive procedures to access ocular fluids. Ocular fluids like aqueous humor and vitreous humor contain valuable molecular information... Read moreTechnology
view channel
Portable Biosensor Diagnoses Psychiatric Disorders Using Saliva Samples
Early diagnosis of psychiatric disorders such as depression, schizophrenia, and bipolar disorder remains one of medicine’s most pressing challenges. Current diagnostic methods rely heavily on clinical... Read more
Cell-Sorting Device Uses Electromagnetic Levitation to Precisely Direct Cell Movement
Sorting different cell types—such as cancerous versus healthy or live versus dead cells—is a critical task in biology and medicine. However, conventional methods often require labeling, chemical exposure,... Read moreIndustry
view channel
Qiagen Acquires Single-Cell Omics Firm Parse Biosciences
QIAGEN (Venlo, Netherlands) has entered into a definitive agreement to fully acquire Parse Biosciences (Seattle, WA, USA), a provider of scalable, instrument-free solutions for single-cell research.... Read more
Puritan Medical Products Showcasing Innovation at AMP2025 in Boston
Puritan Medical Products (Guilford, ME, USA), the world’s most trusted manufacturer of swabs and specimen collection devices, is set to exhibit at AMP2025 in Boston, Massachusetts, from November 11–15.... Read more
Advanced Instruments Merged Under Nova Biomedical Name
Advanced Instruments (Norwood, MA, USA) and Nova Biomedical (Waltham, MA, USA) are now officially doing business under a single, unified brand. This transformation is expected to deliver greater value... Read more