Elecsys Androstenedione Immunoassay Evaluated and Compared
By LabMedica International staff writers Posted on 20 May 2022 |
Androstenedione (ASD) is a biomarker used in the diagnostic workup of hyperandrogenism, congenital adrenal hyperplasia, premature adrenarche, and polycystic ovary syndrome (PCOS). Androstenedione is the precursor of testosterone and estrone and is produced in the gonads and adrenal glands.
Along with other sex hormones, ASD is essential for the onset and development of sexual differentiation in males and females. ASD measurement in serum or plasma is used to assess adrenal and gonadal function and steroid hormone biosynthesis deficiency. A new ASD competitive electrochemiluminescence immunoassay has been evaluated.
Medical Laboratory Scientists at the Cleveland Clinic (Cleveland, OH, USA) evaluated the linearity/analytical measuring range (AMR), precision, and accuracy of the Elecsys ASD assay on the cobas e601 analyzer (Roche Diagnostics, Indianapolis, IN, USA). For comparison, the Immulite ASD was performed using the Immulite 2000 (Siemens Medical Solutions USA, Inc. Malvern, PA, USA). Specimens were sent to a reference laboratory for analysis by LC-MS/MS. The scientists used 40 specimens (33 sera and seven plasma samples) for the method comparison studies The Elecsys ASD assay is an 18-minute competitive electrochemiluminescence immunoassay that utilizes ASD-specific biotinylated monoclonal antibodies and ruthenium complex-conjugated ASD for ASD assessment.
The investigators reported that the Elecsys ASD assay had a linear response across the AMR. The intra- and inter-assay coefficients of variation at various concentrations were ≤4.5%. The Elecsys ASD assay had a mean difference of −0.04 ng/mL (−1.7%) with the LC-MS/MS assay, whereas the Immulite assay had a mean difference of 1.17 ng/mL (66%) and −1.22 ng/mL (−38%) compared to the LC-MS/MS and Elecsys ASD assays, respectively. The Roche recommended RIs for healthy men (0.280–1.52 ng/mL) and postmenopausal women (0.187–1.07 ng/mL) were successfully verified. The RIs for children were adopted from published data. For pre-menopausal women, a RI of <1.60 ng/mL was established. The ASD concentrations in women with and without PCOS overlapped.
The authors concluded that the Elecsys ASD assay has excellent precision, specificity, and accuracy. It offers many advantages over the Immulite ASD assay and is a suitable alternative for ASD measurement in clinical laboratories. However, the assay should be thoroughly validated before its use for clinical testing and reference intervals established for the patient population, as needed. The Elecsys ASD assay has superior comparability to the LC-MS/MS assay than the Immulite assay. The study was published on May 10, 2022 in the journal Practical Laboratory Medicine.
Related Links:
Cleveland Clinic
Roche Diagnostics
Siemens Medical Solutions USA
Latest Technology News
- New Diagnostic System Achieves PCR Testing Accuracy
- DNA Biosensor Enables Early Diagnosis of Cervical Cancer
- Self-Heating Microfluidic Devices Can Detect Diseases in Tiny Blood or Fluid Samples
- Breakthrough in Diagnostic Technology Could Make On-The-Spot Testing Widely Accessible
- First of Its Kind Technology Detects Glucose in Human Saliva
- Electrochemical Device Identifies People at Higher Risk for Osteoporosis Using Single Blood Drop
- Novel Noninvasive Test Detects Malaria Infection without Blood Sample
- Portable Optofluidic Sensing Devices Could Simultaneously Perform Variety of Medical Tests
- Point-of-Care Software Solution Helps Manage Disparate POCT Scenarios across Patient Testing Locations
- Electronic Biosensor Detects Biomarkers in Whole Blood Samples without Addition of Reagents
- Breakthrough Test Detects Biological Markers Related to Wider Variety of Cancers
- Rapid POC Sensing Kit to Determine Gut Health from Blood Serum and Stool Samples
- Device Converts Smartphone into Fluorescence Microscope for Just USD 50
- Wi-Fi Enabled Handheld Tube Reader Designed for Easy Portability