Novel Fecal Calprotectin Test Validated in Pediatric Population
By LabMedica International staff writers Posted on 10 Nov 2020 |
Image: The Buhlmann fCal Turbo turbidimetric immunoassay measures fecal calprotectin in stool specimens (Photo courtesy of Buhlmann Diagnostics).
Inflammatory bowel disease (IBD) and irritable bowel syndrome (IBS) are increasingly prevalent disorders with overlapping clinical presentations. Distinguishing between these two classes of disease is important, as while IBS is an idiopathic functional disorder, IBD is characterized by chronic inflammation and tissue destruction.
Fecal calprotectin (FC) is a biomarker that has shown promise for identification and monitoring of IBD. Calprotectin is a calcium- and zinc-binding protein that is abundant in polynuclear neutrophils, monocytes, and macrophages. Its concentration in stool is well correlated with disease activity of IBD, especially in respect to the colonic mucosa, as measured by histopathological examination.
Medical Scientists at Baylor College of Medicine (Houston, TX, USA) and their colleagues evaluated the Buhlmann fCal Turbo turbidimetric immunoassay (Buhlmann Diagnostics, Amherst, NH, USA) for use on the Vitros 5600 automated chemistry analyzer (Ortho Clinical Diagnostics, Raritan, NJ, USA). The fCal Turbo assay yields measurements in μg/g. Samples used for validation were extracted from stool using the Buhlmann Diagnostics’ Calex cap extraction device. Method comparison was performed with 20 fecal samples with the Buhlmann fCal ELISA, and reference range verification was performed with 33 fecal samples.
The scientists reported that the fCal Turbo assay on the Vitros 5600 was linear between 33.1 and 14,182.5 μg/g, with dilution studies extending the range to 33.1 to 22,000 μg/g, Reproducibility of the assay met acceptability criteria, with intra-assay CV of 0.3–3.2% and inter-assay CV of 5.2–8.9%. Interference studies identified acceptable thresholds for protein, bilirubin, and lipids. They verified a reference range of 33.1 to 60 μg/g in their pediatric patient population. Deming regression identified acceptable correlation with minor positive bias (2.7%) between the fCal Turbo and fCal ELISA methods. In their laboratory, the process from receipt to result was accomplished in approximately 30 minutes.
The authors concluded that the newly FDA-cleared fCal Turbo assay, with the Calex cap for sample extraction, represented a much more rapid and facile workflow for their laboratory than referring samples for analysis by ELISA. They determined that the fCal Turbo assay on the Vitros 5600 analyzer offered excellent performance relative to ELISA, was reproducible, and was not excessively prone to analytical interference. They anticipate that it will be a useful tool for identification of IBD in our patient population. The study was published on October 17, 2020 in the journal Practical Laboratory Medicine.
Related Links:
Baylor College of Medicine
Buhlmann Diagnostics
Ortho Clinical Diagnostics
Fecal calprotectin (FC) is a biomarker that has shown promise for identification and monitoring of IBD. Calprotectin is a calcium- and zinc-binding protein that is abundant in polynuclear neutrophils, monocytes, and macrophages. Its concentration in stool is well correlated with disease activity of IBD, especially in respect to the colonic mucosa, as measured by histopathological examination.
Medical Scientists at Baylor College of Medicine (Houston, TX, USA) and their colleagues evaluated the Buhlmann fCal Turbo turbidimetric immunoassay (Buhlmann Diagnostics, Amherst, NH, USA) for use on the Vitros 5600 automated chemistry analyzer (Ortho Clinical Diagnostics, Raritan, NJ, USA). The fCal Turbo assay yields measurements in μg/g. Samples used for validation were extracted from stool using the Buhlmann Diagnostics’ Calex cap extraction device. Method comparison was performed with 20 fecal samples with the Buhlmann fCal ELISA, and reference range verification was performed with 33 fecal samples.
The scientists reported that the fCal Turbo assay on the Vitros 5600 was linear between 33.1 and 14,182.5 μg/g, with dilution studies extending the range to 33.1 to 22,000 μg/g, Reproducibility of the assay met acceptability criteria, with intra-assay CV of 0.3–3.2% and inter-assay CV of 5.2–8.9%. Interference studies identified acceptable thresholds for protein, bilirubin, and lipids. They verified a reference range of 33.1 to 60 μg/g in their pediatric patient population. Deming regression identified acceptable correlation with minor positive bias (2.7%) between the fCal Turbo and fCal ELISA methods. In their laboratory, the process from receipt to result was accomplished in approximately 30 minutes.
The authors concluded that the newly FDA-cleared fCal Turbo assay, with the Calex cap for sample extraction, represented a much more rapid and facile workflow for their laboratory than referring samples for analysis by ELISA. They determined that the fCal Turbo assay on the Vitros 5600 analyzer offered excellent performance relative to ELISA, was reproducible, and was not excessively prone to analytical interference. They anticipate that it will be a useful tool for identification of IBD in our patient population. The study was published on October 17, 2020 in the journal Practical Laboratory Medicine.
Related Links:
Baylor College of Medicine
Buhlmann Diagnostics
Ortho Clinical Diagnostics
Latest Immunology News
- Diagnostic Blood Test for Cellular Rejection after Organ Transplant Could Replace Surgical Biopsies
- AI Tool Precisely Matches Cancer Drugs to Patients Using Information from Each Tumor Cell
- Genetic Testing Combined With Personalized Drug Screening On Tumor Samples to Revolutionize Cancer Treatment
- Testing Method Could Help More Patients Receive Right Cancer Treatment
- Groundbreaking Test Monitors Radiation Therapy Toxicity in Cancer Patients
- State-Of-The Art Techniques to Investigate Immune Response in Deadly Strep A Infections
- Novel Immunoassays Enable Early Diagnosis of Antiphospholipid Syndrome
- New Test Could Predict Immunotherapy Success for Broader Range Of Cancers
- Simple Blood Protein Tests Predict CAR T Outcomes for Lymphoma Patients
- Cell Sorter Chip Technology to Pave Way for Immune Profiling at POC
- Chip Monitors Cancer Cells in Blood Samples to Assess Treatment Effectiveness
- Automated Immunohematology Approaches Can Resolve Transplant Incompatibility
- AI Leverages Tumor Genetics to Predict Patient Response to Chemotherapy
- World’s First Portable, Non-Invasive WBC Monitoring Device to Eliminate Need for Blood Draw
- Predictive T-Cell Test Detects Immune Response to Viruses Even Before Antibodies Form
- Single Blood Draw to Detect Immune Cells Present Months before Flu Infection Can Predict Symptoms