Automated Hematology Analyzer Assessed for Malaria Diagnosis
By LabMedica International staff writers Posted on 11 Jun 2019 |
Image: The XN-30 automated hematology analyzer that has an automatic measurement functions for red blood cells infected by malaria parasites (Photo courtesy of Sysmex).
Malaria remains a major cause of morbidity and mortality around the world. Sub-Saharan Africa is most affected with Plasmodium falciparum accounting for 99% of estimated cases. Timely and accurate diagnosis of malaria is essential for disease management and control.
Thin and thick blood smear microscopy and malaria rapid diagnostic tests (RDTs) are standard malaria diagnostics in endemic areas. RDTs have significantly improved the use of diagnostics for malaria diagnosis, accounting for 74% of diagnostic tests performed among suspected cases in 2015. Both techniques provide challenges in clinical practice.
An international team of scientists led by Radboud University Medical Centre (Nijmegen, the Netherlands) performed a prospective, double-blinded, phase 3 diagnostic accuracy study on 16 healthy, malaria-naive controlled human malaria infection (CHMI) participants were challenged with five P. falciparum-infected mosquitoes and a diagnostic accuracy study in Burkina Faso. Blood was sampled daily for XN-30, blood smear microscopy, and malaria quantitative polymerase chain reaction (qPCR). The XN-30 is a novel automated hematology analyzer and malaria diagnostic that directly detects and quantifies Plasmodium parasites (falciparum and non-falciparum) in blood using violet laser technology.
The team reported that all CHMI participants became parasitemic by qPCR and XN-30 with a strong correlation for parasite density. The XN-30 accurately monitored treatment and allowed detection of recrudescence. Out of 908 patients in the accuracy study, 241 had microscopic malaria (density 24–491,802 parasites/μL). The sensitivity and specificity of XN-30 compared to microscopy were 98.7% and 99.4%. Results were corrected for qPCR-confirmed sub-microscopic cases. Three microscopy-confirmed cases were not detected by XN-30. However, XN-30 detected 19/134 (14.2%) qPCR-confirmed cases missed by microscopy.
The authors concluded that the XN-30 holds promise as a rapid and sensitive test for malaria detection and subsequent treatment monitoring. Since XN-30 provides a complete blood count (CBC) with each analysis, it provides critical information for malaria management at a price comparable to CBC. Its higher sensitivity compared to microscopy in low-density parasitaemia also makes it a useful tool for mass screening in control programs. The study was published on May 31, 2019, in the journal BMC Medicine.
Related Links:
Radboud University Medical Centre
Thin and thick blood smear microscopy and malaria rapid diagnostic tests (RDTs) are standard malaria diagnostics in endemic areas. RDTs have significantly improved the use of diagnostics for malaria diagnosis, accounting for 74% of diagnostic tests performed among suspected cases in 2015. Both techniques provide challenges in clinical practice.
An international team of scientists led by Radboud University Medical Centre (Nijmegen, the Netherlands) performed a prospective, double-blinded, phase 3 diagnostic accuracy study on 16 healthy, malaria-naive controlled human malaria infection (CHMI) participants were challenged with five P. falciparum-infected mosquitoes and a diagnostic accuracy study in Burkina Faso. Blood was sampled daily for XN-30, blood smear microscopy, and malaria quantitative polymerase chain reaction (qPCR). The XN-30 is a novel automated hematology analyzer and malaria diagnostic that directly detects and quantifies Plasmodium parasites (falciparum and non-falciparum) in blood using violet laser technology.
The team reported that all CHMI participants became parasitemic by qPCR and XN-30 with a strong correlation for parasite density. The XN-30 accurately monitored treatment and allowed detection of recrudescence. Out of 908 patients in the accuracy study, 241 had microscopic malaria (density 24–491,802 parasites/μL). The sensitivity and specificity of XN-30 compared to microscopy were 98.7% and 99.4%. Results were corrected for qPCR-confirmed sub-microscopic cases. Three microscopy-confirmed cases were not detected by XN-30. However, XN-30 detected 19/134 (14.2%) qPCR-confirmed cases missed by microscopy.
The authors concluded that the XN-30 holds promise as a rapid and sensitive test for malaria detection and subsequent treatment monitoring. Since XN-30 provides a complete blood count (CBC) with each analysis, it provides critical information for malaria management at a price comparable to CBC. Its higher sensitivity compared to microscopy in low-density parasitaemia also makes it a useful tool for mass screening in control programs. The study was published on May 31, 2019, in the journal BMC Medicine.
Related Links:
Radboud University Medical Centre
Latest Microbiology News
- Enhanced Rapid Syndromic Molecular Diagnostic Solution Detects Broad Range of Infectious Diseases
- Clinical Decision Support Software a Game-Changer in Antimicrobial Resistance Battle
- New CE-Marked Hepatitis Assays to Help Diagnose Infections Earlier
- 1 Hour, Direct-From-Blood Multiplex PCR Test Identifies 95% of Sepsis-Causing Pathogens
- Mouth Bacteria Test Could Predict Colon Cancer Progression
- Unique Metabolic Signature Could Enable Sepsis Diagnosis within One Hour of Blood Collection
- Groundbreaking Diagnostic Platform Provides AST Results With Unprecedented Speed
- Simple Blood Test Combined With Personalized Risk Model Improves Sepsis Diagnosis
- Blood Analysis Predicts Sepsis and Organ Failure in Children
- TB Blood Test Could Detect Millions of Silent Spreaders
- New Blood Test Cuts Diagnosis Time for Nontuberculous Mycobacteria Infections from Months to Hours
- New Tuberculosis Test to Expand Testing Access in Low- and Middle-Income Countries
- Rapid Test Diagnoses Tropical Disease within Hours for Faster Antibiotics Treatment
- Rapid Molecular Testing Enables Faster, More Targeted Antibiotic Treatment for Pneumonia
- Rapid AST Platform Provides Targeted Therapeutic Results Days Faster Than Current Standard of Care
- New Analysis Method Detects Pathogens in Blood Faster and More Accurately by Melting DNA