Phoenix CPO Test Evaluated for Detection of Carbapenemase Producers

By LabMedica International staff writers
Posted on 14 Nov 2019
The use of nucleic acid amplification tests (NAAT) to identify rapidly Carbapenemase-Producing Organisms (CPO) from samples such as positive blood culture and stools is an alternative approach to establish rapidly an optimal empirical antibiotic therapy or for infectious control management.

Most carbapenemase NAAT panels detect the most frequent gene variants encoding for carbapenemases including KPC, NDM, OXA-48, VIM and IMP. However, these tests cannot exclude the presence of a gene encoding another carbapenemase type and are relatively expensive, hence preventing their use to screen systematically some specimen types or bacterial isolates with CPO suspicion, especially in low to very low prevalence countries.

Image: The BD Phoenix CPO detect test will allow hospitals to identify infections caused by carbapenemase-producing organisms (CPOs) (Photo courtesy of Becton-Dickinson).
Image: The BD Phoenix CPO detect test will allow hospitals to identify infections caused by carbapenemase-producing organisms (CPOs) (Photo courtesy of Becton-Dickinson).

Medical microbiologists at the Lausanne University Hospital (Lausanne, Switzerland) tested the performance of the Phoenix CPO Detect Test (Becton-Dickinson, Franklin Lakes, NJ, USA) on a collection of 92 molecularly-characterized CPO and 93 non-CPO. The collection included 27 non-fermentative bacteria and 158 Enterobacterales. In addition, 295 clinical isolates, including 135 isolates with suspected carbapenemase, were prospectively, but not consecutively isolated from various de-identified clinical specimens during a 10 months period ranging from January to October 2018.

MALDI-TOF was used as an additional phenotypic test to exclude the presence of carbapenemase activity in Phoenix CPO Detect test false positives. All false positive results of the Phoenix CPO Detect test from the prospective study were verified by NAAT testing using the BD MAX Check-Points CPO test. Whole genome sequencing and analysis with libraries prepared with the Nextera XT kit (Illumina, San Diego, USA).

In a retrospective study the scientists reported the Phoenix CPO test exhibited 92.4% accuracy, 97.8% sensitivity 87.1% specificity for carbapenemase detection. The Phoenix CPO test provided a classification to class A, B, and D for 81.3% of detected carbapenemases with 94.6% accuracy. In a prospective study, the CPO test detection performance showed 77.8% accuracy, 100% sensitivity and 67.8% specificity on 135 CPO suspicious isolates and 98.8% accuracy and specificity on 160 non-CPO suspicious isolates. Compared to routine testing, the implementation of the CPO test allowed a mean reduction of 21.3 hours in turnaround time (TAT), 16.8 minutes) in hands-on time (HOT), and CHF 20.6 in costs.

The authors concluded that the Phoenix CPO Detect test likely represents a new diagnostic tool with added value for the detection and management of CPO infection and colonization. The CPO test is reliable for the detection of CPO with a high sensitivity but the relatively low specificity requires the use of additional confirmatory methods. The carbapenemase classification accuracy is robust to provide preliminary results before molecular characterization. The study was published on October 18, 2019 in the journal Clinical Microbiology and Infection.

Related Links:
Lausanne University Hospital
Becton-Dickinson
Illumina



Platinum Member
COVID-19 Rapid Test
OSOM COVID-19 Antigen Rapid Test
Magnetic Bead Separation Modules
MAG and HEATMAG
Anti-Cyclic Citrullinated Peptide Test
GPP-100 Anti-CCP Kit
Gold Member
Fully Automated Cell Density/Viability Analyzer
BioProfile FAST CDV

Latest Microbiology News

Mouth Bacteria Test Could Predict Colon Cancer Progression

Unique Metabolic Signature Could Enable Sepsis Diagnosis within One Hour of Blood Collection

Groundbreaking Diagnostic Platform Provides AST Results With Unprecedented Speed