Elecsys Androstenedione Immunoassay Evaluated and Compared

By LabMedica International staff writers
Posted on 20 May 2022

Androstenedione (ASD) is a biomarker used in the diagnostic workup of hyperandrogenism, congenital adrenal hyperplasia, premature adrenarche, and polycystic ovary syndrome (PCOS). Androstenedione is the precursor of testosterone and estrone and is produced in the gonads and adrenal glands.

Along with other sex hormones, ASD is essential for the onset and development of sexual differentiation in males and females. ASD measurement in serum or plasma is used to assess adrenal and gonadal function and steroid hormone biosynthesis deficiency. A new ASD competitive electrochemiluminescence immunoassay has been evaluated.


Image: The new Elecsys Androstenedione assay has been compared and evaluated (Photo courtesy of Blood Tests London)
Image: The new Elecsys Androstenedione assay has been compared and evaluated (Photo courtesy of Blood Tests London)

Medical Laboratory Scientists at the Cleveland Clinic (Cleveland, OH, USA) evaluated the linearity/analytical measuring range (AMR), precision, and accuracy of the Elecsys ASD assay on the cobas e601 analyzer (Roche Diagnostics, Indianapolis, IN, USA). For comparison, the Immulite ASD was performed using the Immulite 2000 (Siemens Medical Solutions USA, Inc. Malvern, PA, USA). Specimens were sent to a reference laboratory for analysis by LC-MS/MS. The scientists used 40 specimens (33 sera and seven plasma samples) for the method comparison studies The Elecsys ASD assay is an 18-minute competitive electrochemiluminescence immunoassay that utilizes ASD-specific biotinylated monoclonal antibodies and ruthenium complex-conjugated ASD for ASD assessment.

The investigators reported that the Elecsys ASD assay had a linear response across the AMR. The intra- and inter-assay coefficients of variation at various concentrations were ≤4.5%. The Elecsys ASD assay had a mean difference of −0.04 ng/mL (−1.7%) with the LC-MS/MS assay, whereas the Immulite assay had a mean difference of 1.17 ng/mL (66%) and −1.22 ng/mL (−38%) compared to the LC-MS/MS and Elecsys ASD assays, respectively. The Roche recommended RIs for healthy men (0.280–1.52 ng/mL) and postmenopausal women (0.187–1.07 ng/mL) were successfully verified. The RIs for children were adopted from published data. For pre-menopausal women, a RI of <1.60 ng/mL was established. The ASD concentrations in women with and without PCOS overlapped.

The authors concluded that the Elecsys ASD assay has excellent precision, specificity, and accuracy. It offers many advantages over the Immulite ASD assay and is a suitable alternative for ASD measurement in clinical laboratories. However, the assay should be thoroughly validated before its use for clinical testing and reference intervals established for the patient population, as needed. The Elecsys ASD assay has superior comparability to the LC-MS/MS assay than the Immulite assay. The study was published on May 10, 2022 in the journal Practical Laboratory Medicine.

Related Links:
Cleveland Clinic 
Roche Diagnostics 
Siemens Medical Solutions USA 


Gold Member
Troponin T QC
Troponin T Quality Control
Antipsychotic TDM AssaysSaladax Antipsychotic Assays
New
Lab Sample Rotator
H5600 Revolver
New
Vitamin B12 Test
CHORUS CLIA VIT B12

Latest Immunology News

Simple Blood Test Could Detect Drug Resistance in Ovarian Cancer Patients

Advanced Imaging Method Maps Immune Cell Connections to Predict Cancer Patients Survival

Computational Tool Predicts Immunotherapy Outcomes for Metastatic Breast Cancer Patients