Two Vitamin D Immunoassays Compared

By LabMedica International staff writers
Posted on 20 Aug 2019
Vitamin D can be synthesized in skin from 7-dehydrocholesterol following ultraviolet light exposure, which produces Vitamin D3 (cholecalciferol). Vitamin D2 (ergocalciferol) is present in plant sources, and can be absorbed from the diet.

Measurement of 25-Hydroxyvitamin D (25-OH-D) is commonly performed to assess Vitamin D status in clinical laboratories; however, the ability of these assays to detect Vitamin D2 (as 25-OH-D2) or Vitamin D3 (as 25-OH-D3) varies. Assessment of Vitamin D status by measurement of 25-Hydroxyvitamin D (25-OH-D) is widely performed by immunoassay.

Image: The Architect 25-OH Vitamin D reagent kit (Photo courtesy of Abbott).
Image: The Architect 25-OH Vitamin D reagent kit (Photo courtesy of Abbott).

Biochemists at Baylor College of Medicine (Houston, TX, USA) and their colleagues evaluated the performance of the institution's total 25-OH-D assay on the Architect i1000 SR platform to detect either 25-OH-D2 or 25-OH-D3. This assay is a chemiluminescent microparticle immunoassay. Samples spiked with either 25-OH-D2 or 25-OH-D3 and reference laboratory samples with either 25-OH-D2 or 25-OH-D3 concentrations determined by liquid chromatography-tandem mass spectrometry (LC-MS/MS) were used for evaluation. As a comparison to the Architect assay, samples were also tested on the Cobas 8000, which uses an electrochemiluminescence binding assay.

The scientists reported that recovery of 25-OH-D3 in spiked samples was similar by Architect (84%–87%) and Cobas (90%). Recovery of 25-OH-D2 was lower than 25-OH-D3, and was poorer by Architect (37%–40%) than by Cobas (69%–71%). In measurement of samples with known 25-OH-D concentrations, performance of Architect and Cobas assays was similar for 25-OH-D3. However, at concentrations greater than 50 nmol/L 25-OH-D2, the Architect assay exhibited large average negative bias of −27%.

The authors concluded that while the Architect and Cobas assays performed similarly in detection of 25-OH-D3, the Architect assay was significantly poorer at detecting 25-OH-D2 than Cobas, with poorer recovery and significant negative bias at higher concentrations of 25-OH-D2. This agrees with other studies, and indicates that caution should be used in interpreting Architect 25-OH-D results in patients supplemented with Vitamin D2. The study was published on July 29, 2019, in the journal Practical Laboratory Medicine.


Gold Member
TORCH Panel Rapid Test
Rapid TORCH Panel Test
Automated Blood Typing System
IH-500 NEXT
New
Toxoplasma Gondii Test
Toxo IgG ELISA Kit
New
Silver Member
Benchtop Image Acquisition Device
Microwell Imager

Latest Clinical Chem. News

POC Saliva Testing Device Predicts Heart Failure in 15 Minutes

Screening Tool Detects Multiple Health Conditions from Single Blood Drop

Integrated Chemistry and Immunoassay Analyzer with Extensive Assay Menu Offers Flexibility, Scalability and Data Commutability